Advertisement
Original Research Article| Volume 26, 100430, April 2023

On the correction of respiratory motion-induced image reconstruction errors in positron-emission tomography-guided radiation therapy

Open AccessPublished:March 15, 2023DOI:https://doi.org/10.1016/j.phro.2023.100430

      Abstract

      Background and purpose

      Free breathing (FB) positron emission tomography (PET) images are routinely used in radiotherapy for lung cancer patients. Respiration-induced artifacts in these images compromise treatment response assessment and obstruct clinical implementation of dose painting and PET-guided radiotherapy. The purpose of this study is to develop a blurry image decomposition (BID) method to correct motion-induced image-reconstruction errors in FB-PETs.

      Materials and methods

      Assuming a blurry PET is represented as an average of multi-phase PETs. A four-dimensional computed-tomography image is deformably registered from the end-inhalation (EI) phase to other phases. With the registration-derived deformation maps, PETs at other phases can be deformed from a PET at the EI phase. To reconstruct the EI-PET, the difference between the blurry PET and the average of the deformed EI-PETs is minimized using a maximum-likelihood expectation–maximization algorithm. The developed method was evaluated with computational and physical phantoms as well as PET/CT images acquired from three patients.

      Results

      The BID method increased the signal-to-noise ratio from 1.88 ± 1.05 to 10.5 ± 3.3 and universal-quality index from 0.72 ± 0.11 to 1.0 for the computational phantoms, and reduced the motion-induced error from 69.9% to 10.9% in the maximum of activity concentration and from 317.5% to 8.7% in the full width at half maximum of the physical PET-phantom. The BID-based corrections increased the maximum standardized-uptake values by 17.7 ± 15.4% and reduced tumor volumes by 12.5 ± 10.4% on average for the three patients.

      Conclusions

      The proposed image-decomposition method reduces respiration-induced errors in PET images and holds potential to improve the quality of radiotherapy for thoracic and abdominal cancer patients.

      Keywords

      1. Introduction

      Radiotherapy applications introduced in recent years have raised expectations of quantitatively accurate positron emission tomography (PET) images [
      • Jimenez-Ortega E.
      • Ureba A.
      • Vargas A.
      • Baeza J.A.
      • Wals-Zurita A.
      • Garcia-Gomez F.J.
      • et al.
      Dose painting by means of monte carlo treatment planning at the voxel level.
      ]. For example, the intensities of PET images in tumor regions need to be accurately determined in order to prescribe the target dose for dose painting; the location and spatial distribution of tumors in longitudinal PET images need to be identified and compared to facilitate adaptive treatment planning [
      • Yan D.
      • Chen S.
      • Krauss D.J.
      • Chen P.Y.
      • Chinnaiyan P.
      • Wilson G.D.
      Tumor voxel dose-response matrix and dose prescription function derived using (18)F-FDG PET/CT images for adaptive dose painting by number.
      ] or cancer retreatment [
      • Sharifi H.
      • Zhang H.
      • Bagher-Ebadian H.
      • Lu W.
      • Ajlouni M.I.
      • Jin J.Y.
      • et al.
      Utilization of a hybrid finite-element based registration method to quantify heterogeneous tumor response for adaptive treatment for lung cancer patients.
      ]. Quantitatively accurate PET images are fundamental to these applications. However, clinical PET images usually are acquired in a static free-breathing (FB) mode [
      • Kesner A.
      • Pan T.
      • Zaidi H.
      Data-driven motion correction will replace motion-tracking devices in molecular imaging-guided radiation therapy treatment planning.
      ]. Respiratory motion could smear lesions and change standardized uptake values (SUVs), resulting in the maximum of SUVs (SUVmax) in tumor regions being mis-calculated [
      • Nehmeh S.A.
      Respiratory motion correction strategies in thoracic PET-CT imaging.
      ]. Besides quantitative errors, target definition and contour delineation of small lesions in stereotactic body radiotherapy (SBRT) could be complicated by motion artifacts and reduced signal-to-noise ratios [
      • Dinges J.
      • Nekolla S.G.
      • Bundschuh R.A.
      Motion artifacts in oncological and cardiac PET imaging.
      ].
      Gating is a widely used technique to reduce respiratory motion-induced errors in reconstructed PETs [
      • Nehmeh S.A.
      • Erdi Y.E.
      • Ling C.C.
      • Rosenzweig K.E.
      • Squire O.D.
      • Braban L.E.
      • et al.
      Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer.
      ]. While it is subject to factors such as irregular breathing, baseline shift or intra-phase motion [
      • Aristophanous M.
      • Berbeco R.I.
      • Killoran J.H.
      • Yap J.T.
      • Sher D.J.
      • Allen A.M.
      • et al.
      Clinical utility of 4D FDG-PET/CT scans in radiation treatment planning.
      ], gated PETs generally provide more accurate information on tumor location and uptake than 3D PETs. In the recent years, deformable model-based 4D reconstruction methods were proposed to reduce PET data acquisition time [
      • Lamare F.
      • Ledesma Carbayo M.J.
      • Cresson T.
      • Kontaxakis G.
      • Santos A.
      • Le Rest C.C.
      • et al.
      List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations.
      ,
      • Kalantari F.
      • Li T.
      • Jin M.
      • Wang J.
      Respiratory motion correction in 4D-PET by simultaneous motion estimation and image reconstruction (SMEIR).
      ] and PET/MR systems were developed that can avoid anatomical variations during sequential PET and CT data acquisitions [
      • Rank C.M.
      • Heusser T.
      • Wetscherek A.
      • Freitag M.T.
      • Sedlaczek O.
      • Schlemmer H.P.
      • et al.
      Respiratory motion compensation for simultaneous PET/MR based on highly undersampled MR data.
      ]. Furthermore, a generalized reconstruction method was developed to correct motion artifacts in the context of multiple coil acquisition [
      • Fayad H.
      • Odille F.
      • Schmidt H.
      • Wurslin C.
      • Kustner T.
      • Feblinger J.
      • et al.
      The use of a generalized reconstruction by inversion of coupled systems (GRICS) approach for generic respiratory motion correction in PET/MR imaging.
      ]. However, these techniques have not been widely used in clinical practice, and PET images acquired for lung cancer patients still suffer from motion induced artifacts [
      • Kesner A.
      • Pan T.
      • Zaidi H.
      Data-driven motion correction will replace motion-tracking devices in molecular imaging-guided radiation therapy treatment planning.
      ]. For example, in a multi-institutional clinical trial for PET-guided adaptive radiotherapy, FB-PETs were acquired for treatment response assessment and four-dimensional computed-tomography images (4DCTs) acquired for plan adaptation [
      • Kong F.M.
      • Ten Haken R.K.
      • Schipper M.
      • Frey K.A.
      • Hayman J.
      • Gross M.
      • et al.
      Effect of Midtreatment PET/CT-adapted radiation therapy with concurrent chemotherapy in patients with locally advanced non-small-cell lung cancer: a phase 2 clinical trial.
      ]. Although some patients had tumor motion observed in their 4DCTs, there is a lack of an effective tool to correct motion-induced SUV errors in their FB-PETs [
      • Nehmeh S.A.
      Respiratory motion correction strategies in thoracic PET-CT imaging.
      ,
      • Catalano O.A.
      • Masch W.R.
      • Catana C.
      • Mahmood U.
      • Sahani D.V.
      • Gee M.S.
      • et al.
      An overview of PET/MR, focused on clinical applications.
      ].
      In this study, we propose a novel blurry image decomposition (BID) method to convert a blurry image into a motion-frozen (MF) image. Specifically, we use 4DCT images to generate a set of deformation maps and then use these maps to decompose the blurry image into multiple phase images. The equations resulting from the decomposition are solved using a maximum-likelihood expectation–maximization (MLEM) algorithm. We will verify the BID method with both computational and physical phantoms and demonstrate its feasibility of reducing motion artifacts in clinical PET and CT images. The developed BID method has the potential to reduce respiration-induced artifacts in FB-PETs [

      Kong FM, Machtay M, Bradley J, Ten Haken R, Xiao Y, Matuszak M, et al. RTOG 1106/ACRIN 6697: Randomized Phase II Trial of Individualized Adaptive Radiotherapy Using During-Treatment FDG-PET/CT and Modern Technology in Locally Advanced Non-Small Cell Lung Cancer (NSCLC). 2013. https://www.acr.org/-/media/ACR/NOINDEX/Research/ACRIN/Legacy-Trials/ACRIN-6697_RTOG1106.pdf.

      ], enhance the accuracy of treatment response assessment [
      • Eisenhauer E.A.
      • Therasse P.
      • Bogaerts J.
      • Schwartz L.H.
      • Sargent D.
      • Ford R.
      • et al.
      New response evaluation criteria in solid tumours: revised RECIST guideline (version 1.1).
      ], and consequently facilitate clinical applications such dose painting and PET-guided radiotherapy [
      • Arnesen M.R.
      • Knudtsen I.S.
      • Rekstad B.L.
      • Eilertsen K.
      • Dale E.
      • Bruheim K.
      • et al.
      Dose painting by numbers in a standard treatment planning system using inverted dose prescription maps.
      ,
      • Trani D.
      • Yaromina A.
      • Dubois L.
      • Granzier M.
      • Peeters S.G.
      • Biemans R.
      • et al.
      Preclinical assessment of efficacy of radiation dose painting based on intratumoral FDG-PET uptake.
      ,
      • Meng X.
      • Kong F.M.
      • Yu J.
      Implementation of hypoxia measurement into lung cancer therapy.
      ].

      2. Materials and methods

      2.1 The blurry image decomposition (BID) method

      Suppose B is a blurry FB-PET image that has been reconstructed from all projection data. To remove respiratory motion-induced artifacts in B, we decompose it into a set of phase images X(k),k=1,,K where K is the number of respiratory phases. Assuming X(k) is deformed from a MF image S with a deformation map fk, as illustrated in Fig. 1, we can use the following procedure to inversely reconstruct S from the blurry image B.
      Figure thumbnail gr1
      Fig. 1A blurry image B is decomposed into multiple phase images X(k) which will be used to reconstruct a motion-frozen image S.
      Let the FB-PET image B be defined by B=[b1,b2,...,bN], and the phase image X(k) defined by X(k)=[x1k,x2k,,xNk], k = 1, …, K, where K is the number of the respiratory phases, and N is the total number of voxels in the image B. Without loss of generality, the MF image S is assumed to be x(1)defined by S=[x11,x21,,xN1]. Suppose a 4DCT dataset is acquired and binned into K phases. A three-dimensional (3D) deformable image registration (DIR) can be performed from phase 1 to phase k, k = 1, …, K, with the resultant deformation map denoted by fk. Then the image X(1) or S can be reconstructed by minimizing the following objective function
      Ω(W,X(1))=i=1,,Nbi-k=1,,Kwkxi(k)2=i=1,,Nbi-k=1,,Kwkxj(k)(1)2
      (1)


      where W={wk} is a set of weighting factors associated with the duration of individual phases. As illustrated in Fig. 1, the mapping function fk maps voxel jk in image X(1) to voxel i in image X(k). Let g(k) denote the index function of the inverse deformation map (fk)-1, then jk=g(k)(i) and consequently
      xj(k)(1)=xg(k)(i)(1)
      (2)


      where g(k)(i)1,2,,N. The derivatives of the objective function Ω with respect to X(1) can be set to zero to generate a set of linear equations
      AXT-BT=0,
      (3)


      where X=[xg(1)(1)1,xg(1)(2)1,,xg(1)(N)1,,xg(K)(1)1,xg(K)(2)1,,xg(K)(N)1], and A is defined by
      A=wkai,j(k)N×(NK),ai,jk=1,ifgki=j0,otherwise
      (4)


      Here A is an asymmetric matrix, consisting of N×(NK) elements with N denoting the total number of voxels in the PET image. ai,jk corresponds to the position (i, j+(k-1) × N) in the matrix A. The element of A at this position is wk when gki=j and zero otherwise. Therefore, there are N variables xi1,i=1,,N, in equation (3) that need to be determined.

      2.2 Implementation of the BID method

      In this study, a maximum likelihood expectation–maximization (MLEM) method [
      • Hudson H.M.
      • Larkin R.S.
      Accelerated image reconstruction using ordered subsets of projection data.
      ] was implemented to solve equation (3). To improve computational efficiency, the matrix A was reformatted with zero entries suppressed, and the resultant matrix was represented as a linked list of those non-zero entries. For a phase-based 4DCT, each respiratory cycle is equally divided into K phases, so the weighting factor for each phase is 1/K. For an amplitude-based 4DCT, the duration of each phase could be slightly different from 1/K. To address this issue, a random function p(μ) is defined by
      p(μ)=1+μ×rand()%200-100/100
      (5)


      where μ is a simulated amplitude and rand () is a function in the C library that returns a pseudo-random integral number. The weighting factor wk can be calculated from p(μ)/K. We may take W = {1K,,1K}as the starting point to find an optimal set W0 and MF image X(1) such that ΩW,X reaches its minimum at the position (W0,X(1)). 4DCT image registrations were performed using an intensity-based, free-form deformable registration algorithm in the MIM software (MIM Software Inc., Cleveland, OH), where the sum of squared differences was used as similarity metric and a modified gradient descent method was used for optimization [
      • Piper J.W.
      • Richmond J.H.
      • Nelson A.S.
      VoxAlign deformation engine.
      ]. A center-of-mass (COM) mapping method was used for image transformation between X(1) and X(k)[
      • Rosu M.
      • Chetty I.J.
      • Balter J.M.
      • Kessler M.L.
      • McShan D.L.
      • Ten Haken R.K.
      Dose reconstruction in deforming lung anatomy: dose grid size effects and clinical implications.
      ].

      2.3 Computational and physical phantoms for verification of the BID method

      Two computational phantoms and one physical phantom were developed to verify the BID method. Details of the computational phantoms were described in the Supplementary Materials (Section S1). The physical phantom used in the experiment was modified from a motion phantom (Anzai medical system, Tokyo) which has a control panel and a cylindrical container connected by a metal rod. A motor was used to drive the container to move sinusoidally in the superior and inferior direction. A capillary tube measuring 40 mm in length and 0.5 mm in inner diameter was affixed to the top of the container, containing approximately 0.2 μCi of fluorine-18 fluorodeoxyglucose (18F-FDG) and orientated perpendicular to the direction of motion. A 4DCT scan of the moving phantom was acquired on a Siemens SOMATOM CT scanner (Siemens Medical Solutions, USA) and sorted into 10 phases at a slice thickness of 2 mm to derive the motion trajectory of the capillary tube.
      The quality of the reconstructed phantom images was evaluated with signal-to-noise ratio (SNR) and universal quality index (UQI). Suppose R and X represent the reference and reconstructed images for the computational phantoms. SNR is defined by SNRX=λXσX-R, where λX is the mean value of the image X and σX-R is the standard deviation of the difference image between R and X [
      • Yan J.
      • Schaefferkoette J.
      • Conti M.
      • Townsend D.
      A method to assess image quality for Low-dose PET: analysis of SNR, CNR, bias and image noise.
      ]. The overall quality of the reconstructed image X was evaluated using UQI [
      • Wang Z.
      • Bovik A.C.
      • Sheikh H.R.
      • Simoncelli E.P.
      Image quality assessment: from error visibility to structural similarity.
      ] which is defined by
      UQIR,X=2λRλX2λRxλR2+λX2σR2+σX2
      (6)


      where θRX=i=1,,NRi-λRXi-λX/N-1 is the joined correlation between the images R and X. The value of UQI is in a range from 0 to 1 and UQI=1 indicates the highest consistency between X and R. For the capillary tube phantom, its static, and moving and BID-reconstructed PET images were evaluated using the maximum of activity concentration (ACmax) and the full width at half maximum (FWHM) of activity profiles along a line in the moving direction, respectively.

      2.4 Acquisition of 4DCT and PET images from patients

      Three lung cancer patients were enrolled in this study under a retrospective protocol approved by the Institutional Review Board of Medical College of Wisconsin (PRO00036446). 4DCT and FB-PET/CT images were acquired from these patients on a SOMATOM CT scanner and a GE Discovery PET/CT scanner (GE Healthcare, Waukesha, WI), respectively. The acquired 4DCT has an in-plane voxel resolution of 1.1 × 1.1 mm2/pixel and a slice thickness of 3.0 mm. The FB-PET images have an in-plane voxel spacing between 2.7 and 5.2 mm and a slice thickness between 3.2 and 4.0 mm. To improve computational efficiency, regions outside patient contours were cropped out from all PET and CT images. The 4DCT images were deformably registered from either the end-of-inhalation (EI) or end-of-exhalation (EE) phase to other phases. The resultant deformation vector fields were assigned to the domain of PET images using a trilinear interpolation method and then employed in the BID method to convert FB-PET images to MF images.

      3. Results

      3.1 Verification of the BID method with the computational phantoms

      The two computational phantoms were used to validate the BID method. The blurred, BID-reconstructed and referenced images were shown in Fig. 2 (a, b, c) and Fig. 2(d–f) for phantoms 1 and 2, respectively. The BID reconstructions were performed with 104 iterations. The difference between the reference and BID-reconstructed images is in the order of O(10−4), as shown in Fig. 2i. The BID reconstructions increased the UQI from 0.72 ± 0.11 to 1.0 and the SNR from 1.88 ± 1.05 to 10.5 ± 3.3 on average for the two computational phantoms.
      Figure thumbnail gr2
      Fig. 2Two computational phantoms: (a-c) the blurry input, BID-reconstructed and reference image for phantom 1; (d-f) the corresponding images for phantom 2; (g) the blurry image imposed with noise (B1N); (h) the image reconstructed from B1N; (i) the image reconstructed from B1 subtracted by R1.
      To test the stability of the BID method, the blurred image of phantom 1 in Fig. 2a was multiplied by the random function p(μ) with μ = 10%, and up to 4 mm random errors were added to the deformation maps of this phantom. The modified blurry image and deformation maps were substituted into the BID method to convert the noisy image to an MF image with the result shown in Fig. 2h. Although compromised by the imaging and displacement errors, the BID reconstruction still increased the SNR from 2.62 to 4.13 and the UQI from 0.79 to 0.92 for this phantom. The pattern of residuals in Fig. 2(i) is related to the MLEM optimization algorithm. Since the phantom’s motion was simulated in one direction, the residuals of the BID correction are observable only in the moving direction. The convergence of this algorithm is illustrated in Table 3 in the supplementary materials (Section S1).

      3.2 Verification of the BID method with the physical motion phantom

      The physical motion phantom with the FDG source tube is shown in Fig. 3 (a1). The 4DCT images of the phantom were rigidly registered within the MIM software to calculate displacements between different phases. The maximum of the calculated displacements between the EI and EE phases is 19.5 mm which is close to the 20 mm motion amplitude preconfigured in the Anzai phantom. Fig. 3 (b1-c1) shows the coronal and sagittal slices of the PET scan acquired from the moving phantom. The capillary tube was blurred in the phantom’s moving direction and the tube at two extreme positions showed higher activity concentration (AC) than at other transitional positions. The AC difference is due to the fact that the phantom traveled shorter distances in EI or EE phases than in any middle or transitional phase. With moving distances measured from the 4DCT phase images, the BID method was used to decompose the blurred PET. The resultant BID image is shown in Fig. 3(b2–c2) that is highly consistent to the static PET image shown in Fig. 3(b3–c3).
      Figure thumbnail gr3
      Fig. 3(a1) the physical motion phantom; (b1) and (c1) coronal and sagittal slices of the moving PET; (b2) and (c2) coronal and sagittal slices of the BID reconstructed PET; (b3) and (c3) coronal and sagittal slices of the static PET; (a2) profiles of activity concentration in the static, moving and BID-reconstructed PETs along the line of x = 30 (voxels) in the phantom’s moving direction.
      The AC profiles in the moving, static and BID reconstructed PET images along a superior-inferior line that passes through the middle point (x = 30) of the tube were illustrated in Fig. 3(a2). ACmax was reduced from 4.02 × 104 Bq/ml in the static PET to 1.33 × 104 Bq/ml in the moving PET. After the BID correction, the ACmax increased to 4.46 × 104 Bq/ml. The FWHM of the profiles in the moving, static and BID-reconstructed images are 23.47 ± 0.49, 5.61 ± 1.14 and 6.1 ± 0.91 mm, respectively. The BID reconstruction reduced the motion-induced error from 69.9% to 10.9% in the ACmax and from 317.5% to 8.7% in the FWHM of the FDG source tube.

      3.3 MF-PET images reconstructed for the three lung cancer patients

      The average 4DCTs of the three patients were fused to their PET-CTs and consequently aligned to their FB-PETs within the MIM software. The average 4DCT and FB-PET with the target contour for the first patient were shown in Fig. 4a and b. The contour was copied to the BID-reconstructed image (Fig. 4c). The tumor in the reconstructed image is lower than the contour, showing that the BID reconstruction corrected the respiration induced error. Due to DIR and setup uncertainties, there are artifacts in the BID-reconstructed image (Fig. 4c). To address this issue, the FB-PET was down-sampled to the resolution of 0.8 × 0.8 × 0.8 cm3/voxel (Fig. 4e) and overlaid with the average 4DCT in Fig. 4d. The BID method was applied to the down-sampled PET to generate a reconstructed image (Fig. 4f) which shows less artifacts than in Fig. 4c. The location of SUVmax in the BID reconstructed PET was found about 0.62 cm lower than that in the blurry PET as shown in Fig. 4(e, f). This number is consistent with the difference of the tumor position between the EI-4DCT and average 4DCT as shown in the Supplementary materials (Section S2). In addition to the location change, SUVmax increased from 18.9 to 24.3 after the BID reconstruction.
      Figure thumbnail gr4
      Fig. 4(a) the average 4DCT image; (b) the 4-mm FB-PET; (c) the BID-reconstructed PET; (d) the average 4DCT overlaid with the down-sampled 8-mm FB-PET; (e) 8-mm FB-PET; (f) the BID-reconstructed 8-mm PET.
      Fig. 5(a, b) shows a tumor in the right lung for patient 2 and the center of the tumor moves about 2.1 cm between the EE and EI phases. After the BID reconstruction, motion artifacts in the FB-PET (Fig. 5c) are corrected, and the tumor position in the BID-reconstructed PET (Fig. 5d) is consistent to that in the EE-4DCT (Fig. 5a). The BID reconstruction increased SUVmax from 17.1 to 21.3 and reduced the tumor volume from 89.6 cm3 to 68.7 cm3. In contrast to patient 2, patient 3 does not show noticeable tumor motion or deformation between the EE and EI 4DCT images in Fig. 5(e, f), and the BID reconstruction reduces the tumor volume slightly from 36.6 cm3 to 35.7 cm3. On average, the BID reconstruction increased SUVmax by 17.7 ± 15.4% and decreased tumor volume by 12.5 ± 10.4% for the three patients.
      Figure thumbnail gr5
      Fig. 5Images from patients 2 and 3 were illustrated in the upper and lower rows, respectively: (a) EE 4DCT, (b) EI 4DCT, (c) FB-PET overlaid on the EE 4DCT, (d) BID-reconstructed PET overlaid on the EE 4DCT; the correspondent images for patient 3 were shown in (e–h).

      4. Discussion

      Image reconstruction programs such as Software for Tomographic Image Reconstruction (STIR) and Customizable and Advanced Software for Tomographic Reconstruction (CASToR) allow list mode data to be combined with a motion model to reconstruct motion-frozen PETs [
      • Thielemans K.
      • Tsoumpas C.
      • Mustafovic S.
      • Beisel T.
      • Aguiar P.
      • Dikaios N.
      • et al.
      STIR: software for tomographic image reconstruction release 2.
      ,
      • Merlin T.
      • Stute S.
      • Benoit D.
      • Bert J.
      • Carlier T.
      • Comtat C.
      • et al.
      CASToR: a generic data organization and processing code framework for multi-modal and multi-dimensional tomographic reconstruction.
      ]. The motion model can be developed from simultaneously or sequentially acquired MR images [
      • Rank C.M.
      • Heusser T.
      • Wetscherek A.
      • Freitag M.T.
      • Sedlaczek O.
      • Schlemmer H.P.
      • et al.
      Respiratory motion compensation for simultaneous PET/MR based on highly undersampled MR data.
      ,
      • Manber R.
      • Thielemans K.
      • Hutton B.F.
      • Barnes A.
      • Ourselin S.
      • Arridge S.
      • et al.
      Practical PET respiratory motion correction in clinical PET/MR.
      ,
      • Kustner T.
      • Schwartz M.
      • Martirosian P.
      • Gatidis S.
      • Seith F.
      • Gilliam C.
      • et al.
      MR-based respiratory and cardiac motion correction for PET imaging.
      ]. However, list-mode data are neither routinely acquired in clinics nor accessible for cancer treatment [
      • Seo Y.
      • Teo B.K.
      • Hadi M.
      • Schreck C.
      • Bacharach S.L.
      • Hasegawa B.H.
      Quantitative accuracy of PET/CT for image-based kinetic analysis.
      ]. In our previous study [
      • Sharifi H.
      • Zhang H.
      • Bagher-Ebadian H.
      • Lu W.
      • Ajlouni M.I.
      • Jin J.Y.
      • et al.
      Utilization of a hybrid finite-element based registration method to quantify heterogeneous tumor response for adaptive treatment for lung cancer patients.
      ], we investigated changes in longitudinal FB-PETs acquired in a multi-institutional clinical trial for PET-guided adaptive radiotherapy [
      • Kong F.M.
      • Ten Haken R.K.
      • Schipper M.
      • Frey K.A.
      • Hayman J.
      • Gross M.
      • et al.
      Effect of Midtreatment PET/CT-adapted radiation therapy with concurrent chemotherapy in patients with locally advanced non-small-cell lung cancer: a phase 2 clinical trial.
      ]. We found that motion artifacts in these images compromise voxel-based response assessments. In this study, we proposed an image decomposition method that employs a deformable model to transform a blurry PET into a motion-frozen one, eliminating the need for list-mode data [
      • Qiao F.
      • Pan T.
      • Clark Jr., J.W.
      • Mawlawi O.
      Joint model of motion and anatomy for PET image reconstruction.
      ,
      • Grimm R.
      • Furst S.
      • Dregely I.
      • Forman C.
      • Hutter J.M.
      • Ziegler S.I.
      • et al.
      Self-gated radial MRI for respiratory motion compensation on hybrid PET/MR systems.
      ,
      • Brown R.
      • Kolbitsch C.
      • Delplancke C.
      • Papoutsellis E.
      • Mayer J.
      • Ovtchinnikov E.
      • et al.
      Motion estimation and correction for simultaneous PET/MR using SIRF and CIL.
      ].
      There are multiple strategies proposed to remove motion artifacts in FB-PETs. For example, a patient-specific deconvolution kernels was developed to remove rigid motion induced artifacts in a local region [
      • El Naqa I.
      • Low D.A.
      • Bradley J.D.
      • Vicic M.
      • Deasy J.O.
      Deblurring of breathing motion artifacts in thoracic PET images by deconvolution methods.
      ]. However, this method is not applicable to largely deformed structures as the kernel is invariant in the local region. The BID method, on the other hand, can be considered as a generalized deconvolution method that uses a different kernel at each voxel and therefore can be applied to deformed anatomies. Different from deconvolution techniques, a straightforward method was proposed by registering an average 4DCT to each of its phase images to build a deformable model [
      • Gianoli C.
      • Riboldi M.
      • Fontana G.
      • Giri M.G.
      • Grigolato D.
      • Ferdeghini M.
      • et al.
      Optimized PET imaging for 4D treatment planning in radiotherapy: the virtual 4D PET strategy.
      ]. This model was then applied to a blurred 3D PET to create a virtual 4D PET. However, it was reported that this method may reduce motion-induced blurs but cannot restore the maximum of the SUVs [
      • Gianoli C.
      • Riboldi M.
      • Fontana G.
      • Giri M.G.
      • Grigolato D.
      • Ferdeghini M.
      • et al.
      Optimized PET imaging for 4D treatment planning in radiotherapy: the virtual 4D PET strategy.
      ].
      To our knowledge, image processing tools for rigid or elastic motion correction have not yet been implemented in CASToR or STIR [
      • Thielemans K.
      • Tsoumpas C.
      • Mustafovic S.
      • Beisel T.
      • Aguiar P.
      • Dikaios N.
      • et al.
      STIR: software for tomographic image reconstruction release 2.
      ,
      • Merlin T.
      • Stute S.
      • Benoit D.
      • Bert J.
      • Carlier T.
      • Comtat C.
      • et al.
      CASToR: a generic data organization and processing code framework for multi-modal and multi-dimensional tomographic reconstruction.
      ]. In this study, we proposed a voxel-based motion correction method that uses a blurry PET and a 4DCT-based deformable model as input to reconstruct a motion-frozen PET. The developed method can help reduce motion-induced artifacts in PET scans and improve the accuracy of SUV measurements. As 4DCTs are routinely acquired in radiation therapy for treatment planning, the use of the BID method to reduce motion-induced SUV errors in PETs does not add any additional cost to radiotherapy patients.

      Declaration of Competing Interest

      The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

      Acknowledgement

      This project was funded in part by the grants R01-EB028324 and R01-EB032680 from National Institute of Biomedical Imaging and Bioengineering, NIH, USA.

      Appendix A. Supplementary data

      The following are the Supplementary data to this article:

      References

        • Jimenez-Ortega E.
        • Ureba A.
        • Vargas A.
        • Baeza J.A.
        • Wals-Zurita A.
        • Garcia-Gomez F.J.
        • et al.
        Dose painting by means of monte carlo treatment planning at the voxel level.
        Phys Med. 2017; 42: 339-344https://doi.org/10.1016/j.ejmp.2017.04.005
        • Yan D.
        • Chen S.
        • Krauss D.J.
        • Chen P.Y.
        • Chinnaiyan P.
        • Wilson G.D.
        Tumor voxel dose-response matrix and dose prescription function derived using (18)F-FDG PET/CT images for adaptive dose painting by number.
        Int J Radiat Oncol Biol Phys. 2019; 104: 207-218https://doi.org/10.1016/j.ijrobp.2019.01.077
        • Sharifi H.
        • Zhang H.
        • Bagher-Ebadian H.
        • Lu W.
        • Ajlouni M.I.
        • Jin J.Y.
        • et al.
        Utilization of a hybrid finite-element based registration method to quantify heterogeneous tumor response for adaptive treatment for lung cancer patients.
        Phys Med Biol. 2018; 63065017https://doi.org/10.1088/1361-6560/aab235
        • Kesner A.
        • Pan T.
        • Zaidi H.
        Data-driven motion correction will replace motion-tracking devices in molecular imaging-guided radiation therapy treatment planning.
        Med Phys. 2018; 45: 3477-3480https://doi.org/10.1002/mp.12928
        • Nehmeh S.A.
        Respiratory motion correction strategies in thoracic PET-CT imaging.
        PET Clin. 2013; 8: 29-36https://doi.org/10.1016/j.cpet.2012.10.004
        • Dinges J.
        • Nekolla S.G.
        • Bundschuh R.A.
        Motion artifacts in oncological and cardiac PET imaging.
        PET Clin. 2013; 8: 1-9https://doi.org/10.1016/j.cpet.2012.10.001
        • Nehmeh S.A.
        • Erdi Y.E.
        • Ling C.C.
        • Rosenzweig K.E.
        • Squire O.D.
        • Braban L.E.
        • et al.
        Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer.
        Med Phys. 2002; 29: 366-371https://doi.org/10.1118/1.1448824
        • Aristophanous M.
        • Berbeco R.I.
        • Killoran J.H.
        • Yap J.T.
        • Sher D.J.
        • Allen A.M.
        • et al.
        Clinical utility of 4D FDG-PET/CT scans in radiation treatment planning.
        Int J Radiat Oncol Biol Phys. 2012; 82: e99-ehttps://doi.org/10.1016/j.ijrobp.2010.12.060
        • Lamare F.
        • Ledesma Carbayo M.J.
        • Cresson T.
        • Kontaxakis G.
        • Santos A.
        • Le Rest C.C.
        • et al.
        List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations.
        Phys Med Biol. 2007; 52: 5187-5204https://doi.org/10.1088/0031-9155/52/17/006
        • Kalantari F.
        • Li T.
        • Jin M.
        • Wang J.
        Respiratory motion correction in 4D-PET by simultaneous motion estimation and image reconstruction (SMEIR).
        Phys Med Biol. 2016; 61: 5639-5661https://doi.org/10.1088/0031-9155/61/15/5639
        • Rank C.M.
        • Heusser T.
        • Wetscherek A.
        • Freitag M.T.
        • Sedlaczek O.
        • Schlemmer H.P.
        • et al.
        Respiratory motion compensation for simultaneous PET/MR based on highly undersampled MR data.
        Med Phys. 2016; 43: 6234https://doi.org/10.1118/1.4966128
        • Fayad H.
        • Odille F.
        • Schmidt H.
        • Wurslin C.
        • Kustner T.
        • Feblinger J.
        • et al.
        The use of a generalized reconstruction by inversion of coupled systems (GRICS) approach for generic respiratory motion correction in PET/MR imaging.
        Phys Med Biol. 2015; 60: 2529-2546https://doi.org/10.1088/0031-9155/60/6/2529
        • Kong F.M.
        • Ten Haken R.K.
        • Schipper M.
        • Frey K.A.
        • Hayman J.
        • Gross M.
        • et al.
        Effect of Midtreatment PET/CT-adapted radiation therapy with concurrent chemotherapy in patients with locally advanced non-small-cell lung cancer: a phase 2 clinical trial.
        JAMA Oncol. 2017; 3: 1358-1365https://doi.org/10.1001/jamaoncol.2017.0982
        • Catalano O.A.
        • Masch W.R.
        • Catana C.
        • Mahmood U.
        • Sahani D.V.
        • Gee M.S.
        • et al.
        An overview of PET/MR, focused on clinical applications.
        Abdom Radiol (NY). 2017; 42: 631-644https://doi.org/10.1007/s00261-016-0894-5
      1. Kong FM, Machtay M, Bradley J, Ten Haken R, Xiao Y, Matuszak M, et al. RTOG 1106/ACRIN 6697: Randomized Phase II Trial of Individualized Adaptive Radiotherapy Using During-Treatment FDG-PET/CT and Modern Technology in Locally Advanced Non-Small Cell Lung Cancer (NSCLC). 2013. https://www.acr.org/-/media/ACR/NOINDEX/Research/ACRIN/Legacy-Trials/ACRIN-6697_RTOG1106.pdf.

        • Eisenhauer E.A.
        • Therasse P.
        • Bogaerts J.
        • Schwartz L.H.
        • Sargent D.
        • Ford R.
        • et al.
        New response evaluation criteria in solid tumours: revised RECIST guideline (version 1.1).
        Eur J Cancer. 2009; 45: 228-247https://doi.org/10.1016/j.ejca.2008.10.026
        • Arnesen M.R.
        • Knudtsen I.S.
        • Rekstad B.L.
        • Eilertsen K.
        • Dale E.
        • Bruheim K.
        • et al.
        Dose painting by numbers in a standard treatment planning system using inverted dose prescription maps.
        Acta Oncol. 2015; 54: 1607-1613https://doi.org/10.3109/0284186X.2015.1061690
        • Trani D.
        • Yaromina A.
        • Dubois L.
        • Granzier M.
        • Peeters S.G.
        • Biemans R.
        • et al.
        Preclinical assessment of efficacy of radiation dose painting based on intratumoral FDG-PET uptake.
        Clin Cancer Res. 2015; 21: 5511-5518https://doi.org/10.1158/1078-0432.CCR-15-0290
        • Meng X.
        • Kong F.M.
        • Yu J.
        Implementation of hypoxia measurement into lung cancer therapy.
        Lung Cancer. 2012; 75: 146-150https://doi.org/10.1016/j.lungcan.2011.09.009
        • Hudson H.M.
        • Larkin R.S.
        Accelerated image reconstruction using ordered subsets of projection data.
        IEEE Trans Med Imaging. 1994; 13: 601-609https://doi.org/10.1109/42.363108
        • Piper J.W.
        • Richmond J.H.
        • Nelson A.S.
        VoxAlign deformation engine.
        MIM Software, 2018
        • Rosu M.
        • Chetty I.J.
        • Balter J.M.
        • Kessler M.L.
        • McShan D.L.
        • Ten Haken R.K.
        Dose reconstruction in deforming lung anatomy: dose grid size effects and clinical implications.
        Med Phys. 2005; 32: 2487-2495https://doi.org/10.1118/1.1949749
        • Yan J.
        • Schaefferkoette J.
        • Conti M.
        • Townsend D.
        A method to assess image quality for Low-dose PET: analysis of SNR, CNR, bias and image noise.
        Cancer Imaging. 2016; 16: 26https://doi.org/10.1186/s40644-016-0086-0
        • Wang Z.
        • Bovik A.C.
        • Sheikh H.R.
        • Simoncelli E.P.
        Image quality assessment: from error visibility to structural similarity.
        IEEE Trans. 2004; 13: 600-612
        • Thielemans K.
        • Tsoumpas C.
        • Mustafovic S.
        • Beisel T.
        • Aguiar P.
        • Dikaios N.
        • et al.
        STIR: software for tomographic image reconstruction release 2.
        Phys Med Biol. 2012; 57: 867-883https://doi.org/10.1088/0031-9155/57/4/867
        • Merlin T.
        • Stute S.
        • Benoit D.
        • Bert J.
        • Carlier T.
        • Comtat C.
        • et al.
        CASToR: a generic data organization and processing code framework for multi-modal and multi-dimensional tomographic reconstruction.
        Phys Med Biol. 2018; 63185005https://doi.org/10.1088/1361-6560/aadac1
        • Manber R.
        • Thielemans K.
        • Hutton B.F.
        • Barnes A.
        • Ourselin S.
        • Arridge S.
        • et al.
        Practical PET respiratory motion correction in clinical PET/MR.
        J Nucl Med. 2015; 56: 890-896https://doi.org/10.2967/jnumed.114.151779
        • Kustner T.
        • Schwartz M.
        • Martirosian P.
        • Gatidis S.
        • Seith F.
        • Gilliam C.
        • et al.
        MR-based respiratory and cardiac motion correction for PET imaging.
        Med Image Anal. 2017; 42: 129-144https://doi.org/10.1016/j.media.2017.08.002
        • Seo Y.
        • Teo B.K.
        • Hadi M.
        • Schreck C.
        • Bacharach S.L.
        • Hasegawa B.H.
        Quantitative accuracy of PET/CT for image-based kinetic analysis.
        Med Phys. 2008; 35: 3086-3089https://doi.org/10.1118/1.2937439
        • Qiao F.
        • Pan T.
        • Clark Jr., J.W.
        • Mawlawi O.
        Joint model of motion and anatomy for PET image reconstruction.
        Med Phys. 2007; 34: 4626-4639https://doi.org/10.1118/1.2804721
        • Grimm R.
        • Furst S.
        • Dregely I.
        • Forman C.
        • Hutter J.M.
        • Ziegler S.I.
        • et al.
        Self-gated radial MRI for respiratory motion compensation on hybrid PET/MR systems.
        Med Image Comput Comput Assist Interv. 2013; 16: 17-24https://doi.org/10.1007/978-3-642-40760-4_3
        • Brown R.
        • Kolbitsch C.
        • Delplancke C.
        • Papoutsellis E.
        • Mayer J.
        • Ovtchinnikov E.
        • et al.
        Motion estimation and correction for simultaneous PET/MR using SIRF and CIL.
        Philos Trans A Math Phys Eng Sci. 2021; 379: 20200208https://doi.org/10.1098/rsta.2020.0208
        • El Naqa I.
        • Low D.A.
        • Bradley J.D.
        • Vicic M.
        • Deasy J.O.
        Deblurring of breathing motion artifacts in thoracic PET images by deconvolution methods.
        Med Phys. 2006; 33: 3587-3600https://doi.org/10.1118/1.2336500
        • Gianoli C.
        • Riboldi M.
        • Fontana G.
        • Giri M.G.
        • Grigolato D.
        • Ferdeghini M.
        • et al.
        Optimized PET imaging for 4D treatment planning in radiotherapy: the virtual 4D PET strategy.
        Technol Cancer Res Treat. 2015; 14: 99-110https://doi.org/10.7785/tcrt.2012.500393