Original Research Article| Volume 25, 100420, January 2023

Assessing the impact of adaptations to the clinical workflow in radiotherapy using transit in vivo dosimetry

Open AccessPublished:January 28, 2023DOI:https://doi.org/10.1016/j.phro.2023.100420

      Highlights

      • Analyzing in vivo dosimetry data can assist in continuous quality improvement.
      • In vivo dosimetry data can be used to assess adaptations to the clinical workflow.
      • Optimizing clinical workflows leads to enhanced in vivo dosimetry results.
      • Large-scale in vivo dosimetry data can indicate general items for improvement.

      Abstract

      Background and Purpose

      Currently in-vivo dosimetry (IVD) is primarily used to identify individual patient errors in radiotherapy. This study investigated possible correlations of observed trends in transit IVD results, with adaptations to the clinical workflow, aiming to demonstrate the possibility of using the bulk data for continuous quality improvement.

      Materials and methods

      In total 84,100 transit IVD measurements were analyzed of all patients treated between 2018 and 2022, divided into four yearly periods. Failed measurements (FM) were divided per pathology and into four categories of causes of failure: technical, planning and positioning problems, and anatomic changes.

      Results

      The number of FM due to patient related problems gradually decreased from 9.5% to 6.6%, 6.1% and 5.6% over the study period. FM attributed to positioning problems decreased from 10.0% to 4.9% in boost breast cancer patients after introduction of extra imaging, from 9.1% to 3.9% in Head&Neck patients following education of radiation therapists on positioning of patients’ shoulders, from 6.1% to 2.8% in breast cancer patients after introduction of ultrahypofractionated breast radiotherapy with daily online pre-treatment imaging and from 11.2% to 4.3% in extremities following introduction of immobilization with calculated couch parameters and a Surface Guided Radiation Therapy solution. FM related to anatomic changes decreased from 10.2% to 4.0% in rectum patients and from 6.7% to 3.3% in prostate patients following more patient education from dieticians.

      Conclusions

      Our study suggests that IVD can be a powerful tool to assess the impact of adaptations to the clinical workflow and its use for continuous quality improvement.

      Keywords

      Abbreviations:

      IVD (In-vivo Dosimetry), FM (Failed Measurements), H&N (Head&Neck), EPID (Electronic Portal Imaging Device), VMAT (Volumetric Modulated Arc Therapy), PSQA (Patient Specific Quality Assurance), HU (Hounsfield Units)

      1. Introduction

      In-vivo dosimetry (IVD) is recommended in radiotherapy to avoid major treatment errors and to improve accuracy, as elucidated in several reports [
      • Derreumaux S.
      • Etard C.
      • Huet C.
      • Trompier F.
      • Clairand I.
      • Bottollier-Depois J.-F.
      • et al.
      Lessons from recent accidents in radiation therapy in France.
      ,
      • Ortiz López P.
      • Cosset J.M.
      • Dunscombe P.
      • Holmberg O.
      • Rosenwald J.C.
      • Pinillos Ashton L.
      • et al.
      ICRP publication 112. A report of preventing accidental exposures from new external beam radiation therapy technologies.
      ,

      International Atomic Energy Agency. Development of procedures for in vivo dosimetry in radiotherapy. Human Health Report No. 8, IAEA, Vienna, 2013. https://www-pub.iaea.org/MTCD/Publications/PDF/Pub1606_web.pdf.

      ,

      International Atomic Energy Agency. Accuracy Requirements and Uncertainties in Radiotherapy, Human Health Series No. 31, IAEA, Vienna, 2016. https://www-pub.iaea.org/MTCD/Publications/PDF/P1679_HH31_web.pdf.

      ,
      • Noel A.
      • Aletti P.
      • Bey P.
      • Malissard L.
      Detection of errors in individual patients in radiotherapy by systematic in vivo dosimetry.
      ]. IVD using Electronic Portal Imaging Device (EPID) images has become routine practice in a growing number of clinics in recent years. Many groups have reported the capability of EPID IVD to detect dosimetric deviations due to multiple sources, as well as the dosimetric advantages of using IVD: the Netherlands Cancer Institute has published multiple reports showing the importance of EPID IVD [
      • Mans A.
      • Wendling M.
      • McDermott L.N.
      • Sonke J.J.
      • Tielenburg R.
      • Vijlbrief R.
      • et al.
      Catching errors with in vivo EPID dosimetry.
      ,
      • Mijnheer B.
      • González P.
      • Olaciregui-Ruiz I.
      • Rozendaal R.
      • van Herk M.
      • Mans A.
      Overview of 3-year experience with large-scale electronic portal imaging device-based 3-dimensional transit dosimetry Pract.
      ,
      • Mijnheer B.
      • Beddar S.
      • Izewska J.
      • Reft C.
      In vivo dosimetry in external beam radiotherapy.
      ,
      • Mijnheer B.
      • Jomehzadeh A.
      • González P.
      • Olaciregui-Ruiz I.
      • Rozendaal R.
      • Shokrani P.
      • et al.
      Error detection during VMAT delivery using EPID-based 3D transit dosimetry.
      ,
      • Mijnheer B.
      • Olaciregui-Ruiz I.
      • Rozendaal R.
      • Spreeuw H.
      • van Herk M.
      • Mans A.
      Current status of 3D EPID-based in vivo dosimetry in The Netherlands Cancer Institute.
      ], other groups investigated the sensitivity and specificity for the detection of errors [
      • Wolfs C.J.A.
      • Brás M.G.
      • Schyns L.E.J.R.
      • Nijsten S.M.J.J.G.
      • van Elmpt W.
      • Scheib S.G.
      • et al.
      Detection of anatomical changes in lung cancer patients with 2D time-integrated, 2D time-resolved and 3D time-integrated portal dosimetry: a simulation study.
      ,
      • Bojechko C.
      • Ford E.C.
      Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations.
      ,
      • Bawazeer O.
      • Sarasanandarajah S.
      • Herath S.
      • Kron T.
      • Deb P.
      Sensitivity of electronic portal imaging device (EPID) based transit dosimetry to detect inter-fraction patient variations.
      ,
      • Bojechko C.
      • Phillps M.
      • Kalt A.
      • Ford E.C.
      A quantification of the effectiveness of EPID dosimetry and software-based plan verification systems in detecting incidents in radiotherapy.
      ,
      • Zhuang A.
      • Olch A.
      Sensitivity study of an automated system for daily patient QA using EPID exit dose images.
      ,
      • Moustakis C.
      • Ebrahimi Tazehmahalleh F.
      • Elsayad K.
      • Fezeu F.
      • Scobioala S.
      A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetry: a step to QA with in vivo EPID dosimetry.
      ,
      • Doolan P.
      • Nikolaou M.
      • Ferentinos K.
      • Anagnostopoulos G.
      Assessment of a commercial EPID dosimetry system to detect radiotherapy treatment errors.
      ]. For a comprehensive literature review on electronic portal imaging for radiotherapy dosimetry the reader is referred to van Elmpt et al. [
      • Van Elmpt W.
      • McDermott L.
      • Nijsten S.
      • Wendling M.
      • Lambin P.
      • Mijnheer B.
      A literature review of electronic portal imaging for radiotherapy dosimetry.
      ] and McCurdy et al. [

      McCurdy B, Greer P, Bedford J. Electronic portal imaging device dosimetry. In: Mijnheer B. Clinical 3D dosimetry in modern radiation therapy. 2017; 169–98. Electronic Portal Imaging Device Dosimetry | 7 | Clinical 3D Dosimetry (taylorfrancis.com).

      ].
      Recent versions of EPID, increased computer power and fully automated EPID-based systems make it feasible to perform dosimetric patient specific quality assurance (PSQA) on a large scale. This extended usage opens possibilities to not only detect individual errors, but to use the data as a tool to evaluate the quality of treatment procedures and hence as an input for continuous quality improvement. In literature there is very limited data on this subject. Celi et al. [
      • Celi S.
      • Costa E.
      • Wessels C.
      • Mazal A.
      • Fourquet A.
      • Francois P.
      EPID based in vivo dosimetry system: clinical experience and results.
      ] performed an analysis of the measurements per linac and energy over a two-year period including a more detailed examination per technique and treatment site over a six-month period. The prostate study revealed that beams and arcs with out-of-tolerance IVD results tended to have more complex modulation and a lower exposure of the points of interest. By analyzing the long-term PSQA data of prostate Volumetric Modulated Arc Therapy (VMAT) plans, Mans et al. [
      • Mans A.
      • Rozendaal R.
      • Janssen T.
      • Damen E.
      • Kaas J.
      • van Mourik A.
      • et al.
      Reduction of systematic dosimetric uncertainties in volumetric modulated arc therapy triggered by patient-specific quality assurance.
      ] revealed an under-dosage gradual increasing to about 2 % in 3 years correlated with plan complexity and coinciding with changes in clinically applied planning techniques. They also presented a procedure to investigate long-term trend analysis of PSQA data to identify any change with time and to find in a systematic way the reasons for it. The method of statistical process control (SPC) could be very useful for assessing results, as described by Fuangrod et al. [
      • Fuangrod T.
      • Greer P.B.
      • Simpson J.
      • Zwan B.J.
      • Middleton R.H.
      A method for evaluating treatment quality using in vivo EPID dosimetry and statistical process control in radiation therapy.
      ] and Esposito et al. [
      • Esposito M.
      • Ghirelli A.
      • Pini S.
      • Alpi P.
      • Barca R.
      • Fondelli S.
      • et al.
      Clinical implementation of 3D in vivo dosimetry for abdominal and pelvic stereotactic treatments.
      ]. Establishing tolerance limits using the SPC technique can be quite labor-intensive though and this method has not been used by our group.
      To our knowledge, none of the groups have evaluated the impact of changes made to the clinical workflow on the results of their IVD measurements. In the current paper the long-term PSQA data of a large cohort of patients with multiple pathologies is analyzed and correlated to adaptations in the clinical workflow. The aim was to demonstrate the capability of EPID IVD to assess the impact of adaptations to the clinical workflow and hence the possibility to use the bulk data of a longer period of time for continuous quality improvement rather than merely the detection of individual errors.

      2. Materials and methods

      The study was conducted at the Iridium Netwerk, a radiotherapy facility in Belgium with 10 linear accelerators: a mix of Varian Clinac, TrueBeam and TrueBeam STX (Varian Oncology Systems, Palo Alto, Ca, USA). Two treatment planning systems were used to create plans: RayStation (RaySearch Laboratories, Stockholm, Sweden) for stereotactic plans and Eclipse (Varian Medical Systems, Palo Alto, CA, USA) for all other types of plans. An automated web-based system was installed early 2017, for both pre-treatment and in-vivo QA based on EPID measurements (PerFRACTION™, part of SunCHECKTM, Sun Nuclear Corporation (SNC), version 1.7 since October 2017, version 2.0 since May 2018, version 2.2 since November 2019, version 3.1 since April 2021). This study included the data of all patients treated between September 2018 and August 2022: DICOM data were pushed to the server, images were actively retrieved, and calculation and analysis occurred automatically in the background. Failing measurements were checked daily by the responsible physicists as well as by the physicians and appropriate actions were taken. The study was approved by the scientific committee and patient consent was waived.
      Transit EPID integrated images were generated standardly the first three days of treatment and weekly thereafter. More measurements were performed sometimes due to technical problems or when patient follow-up was needed. The parameters for 2D gamma analysis depended on the type of treatment. The templates for this were designed to make a distinction in tolerance levels depending on the treatment site. These tolerances were empirically determined as described in a previous publication from our group [
      • Bossuyt E.
      • Weytjens R.
      • Nevens D.
      • De Vos S.
      • Verellen D.
      Evaluation of automated pre-treatment and transit in-vivo dosimetry in radiotherapy using empirically determined parameters.
      ] and have not been altered since, offering the advantage that results could be safely evaluated and compared over the years. An overview of the parameters for gamma analysis can be found in Table 1.
      Table 1Summary of empirically determined parameters for gamma analysis of in-vivo transit dosimetry results.
      Normalization (Local/Global)Dose Difference Tolerance (%)Distance Tolerance (mm)Low Dose Threshold (%)Passing Tolerance Level (%)
      BreastLocal762090
      Whole Brain RadioTherapyLocal732090
      Palliative treatmentsLocal752093
      H&N and BrainGlobal332095
      RectumGlobal552093
      Other treatment sites with maskGlobal532095
      Other treatment sites without mask (including lung, pelvis, abdomen,…)Global552095
      Stereotactic 1 mmLocal1012095
      Stereotactic 2 mmLocal1022095
      Stereotactic 3 mmLocal1032095
      Results are reported for 84,100 transit IVD measurements of patients treated between September 2018 and August 2022, divided into four yearly periods. Failed measurements (FM) were divided per pathology and into four categories of causes of failure to assess the influence of adaptations to the clinical workflow: technical, planning and positioning problems, and anatomic changes. The categories of causes of failure have been assigned using the comments and notes that were made by the physicists and physicians for each failing measurement. If failure was due to more than 1 reason, the most contributing cause, according to the physicist, was assigned. Technical problems included software bugs, wrong imager position, problems with imager calibration and interrupted beams causing missing dose in the image. Planning problems included errors in body contouring, skin flash tool planning, Hounsfield units (HU) assignment, CT’s with artefacts or wrong HU used for planning. Positioning problems included amongst others shoulder positioning, arm positioning, issues related to the breath hold technique and problems with immobilization devices. Examples of anatomy changes included weight loss, tumor growth or shrinkage, breast swelling and pneumonia.
      Some adaptations to the clinical workflow were investigated in detail: the introduction of extra imaging for the boost in breast cancer after the first year, the education of radiation therapists (RTT’s) on positioning of patients’ shoulders in head & neck cancer (H&N) after the first year, more patient education from dieticians for prostate-, rectum-, upper abdomen- and esophageal cancer patients after the first year, ultrahypofractionated breast RT in five fractions with daily online pre-treatment imaging replacing a 15-fraction scheme in the second year and immobilization with calculated couch parameters and a Surface Guided Radiation Therapy (SGRT) solution (C-RAD) in the third year.

      3. Results

      The number of failed measurements (FM) gradually decreased over the years with 15.7 %, 13.3 %, 10.5 % and 9.0 % of FM in the first, second, third and fourth year respectively of which 6.2 %, 6.8 %, 4.5 % and 3.4 % were technical problems and 9.5 %, 6.5 %, 6.0 % and 5.6 % were caused by patient related issues, which were subdivided in planning problems, positioning problems and anatomic changes (Fig. 1).
      Figure thumbnail gr1
      Fig. 1Overview of the number of all failed measurements over the years, subdivided into four categories of causes of failure: anatomic changes, positioning problems, planning problems and technical problems.
      Table 2 shows the results of the measurements for the investigated pathologies. Relevant numbers related to the investigated adaptations to the clinical workflow are highlighted.
      Table 2Overview of the results of the measurements for prostate, rectum, upper abdomen, head & neck (H&N), extremities, esophagus, breast and boost breast patients, divided into four categories of causes of failure: technical, planning and positioning problems, and anatomic changes. Relevant numbers related to the investigated adaptations to the clinical workflow are highlighted in different colors: extra imaging for the boost in breast cancer in yellow, the education of radiation therapists on positioning of patients’ shoulders in green, patient education from dieticians in blue, ultrahypofractionated breast RT in grey, immobilization with calculated couch parameters and a Surface Guided Radiation Therapy solution in pink.
      FM attributed to positioning problems in boost breast cancer patients decreased from 10.0 % in the first year to 4.9 % in the second year after the introduction of extra imaging. FM attributed to positioning problems in H&N patients decreased from 9.1 % in the first year to 3.9 % in the second year following education of RTT’s on positioning of patients’ shoulders. FM related to anatomic changes in rectum and prostate patients have been reduced from 10.2 % and 6.7 % in the first year to 4.0 % and 3.3 % in the second year respectively, following patient education from dieticians. No difference was observed for stomach and esophageal cancer patients.
      FM attributed to positioning problems in breast cancer patients decreased from 6.1 % in the second year to 2.8 % in the third year after the introduction of ultrahypofractionated breast radiotherapy with daily online pre-treatment imaging. FM attributed to patient positioning decreased further in the fourth year for most pathologies following the introduction of immobilization with calculated couch parameters and an SGRT solution, with the largest effect seen for extremities where FM due to patient positioning decreased from 11.2 % in the third year to 4.3 % in the fourth year.

      4. Discussion

      This study investigated if the observed trends in the IVD results over the years could be a result of adaptations to the clinical workflow, or the other way around, if the impact of adaptations could be monitored by IVD. Our results suggest EPID IVD was indeed able to assess the impact of some adaptations to the clinical workflow and can therefore assist in continuous quality improvement.
      The most relevant observations of the overview of number of FM over the years (Fig. 1) were that the number of FM due to patient related problems gradually decreased from 9.5 % in the first year to 6.6 % in the second year, 6.1 % in the third year and 5.6 % in the fourth year. Between the first and the second year, there was a decrease of 1.9 % in positioning problems and 0.6 % in anatomic changes. In the third year the decrease was mainly caused by a decrease in technical problems and in patient positioning. The number of planning problems and anatomic changes were about the same. The fourth year showed a further small decrease. The increase of FM due to technical problems in the second year was mainly due to a bug in a new version of the IVD software causing a wrong predicted dose for plans with multiple energies. This bug was solved by the third year, so most of the software problems disappeared. The further reduction of technical problems over the years was mainly due to the gradual replacement of older linear accelerators by new machines. The latter experience less imager breakdowns and are equipped with a more stable type of imagers, the imager position of the aS1000 imagers not being corrected mechanically during VMAT and hence less stable than the aS1200 imagers.
      Most studies found in literature analyzing IVD results on a large scale, report deviations of the same type of causes and order of magnitude [
      • Noel A.
      • Aletti P.
      • Bey P.
      • Malissard L.
      Detection of errors in individual patients in radiotherapy by systematic in vivo dosimetry.
      ,
      • Fuangrod T.
      • Greer P.B.
      • Simpson J.
      • Zwan B.J.
      • Middleton R.H.
      A method for evaluating treatment quality using in vivo EPID dosimetry and statistical process control in radiation therapy.
      ,
      • Olch A.J.
      • O’Meara K.
      • Wong K.K.
      First report of the clinical use of a commercial automated system for daily patient QA using EPID exit images.
      ,
      • Esposito M.
      • Piermattei A.
      • Bresciani S.
      • Orlandini L.
      • Falco M.
      • Giancaterino S.
      • et al.
      Improving dose delivery accuracy with EPID in vivo dosimetry: results from a multicenter study.
      ,
      • Nailon W.H.
      • Welsh D.
      • McDonald K.
      • Burns D.
      • Forsyth J.
      • Cooke G.
      • et al.
      EPID-based in vivo dosimetry using Dosimetry Check™: overview and clinical experience in a 5-yr study including breast, lung, prostate, and head and neck cancer patients.
      ] but none to our knowledge, have evaluated the impact of changes made to the clinical workflow on the results of their IVD measurements.
      After the first year, analysis of our results showed a large number of positioning issues for extremities, boost breast patients and Head & Neck (H&N) patients. Therefore, some changes to the clinical workflow were introduced: after the first year extra imaging was applied for boost breast and extra education was given to the RTT’s to pay attention to proper positioning of the shoulders. After the third year immobilization with calculated couch parameters was introduced and an SGRT solution was installed on all treatment machines aiming at reducing the number of failed measurements due to positioning issues, especially for extremities.
      To evaluate the impact of these changes in the workflow, FM of these pathologies were analyzed in more detail. FM attributed to positioning problems in boost breast cancer patients decreased from 10.0 % in the first year to 4.9 % in the second year (highlighted in yellow in Table 2). The large number of technical problems in the second year was caused by a software bug where the dose prediction of plans with multiple energies, which is often used for boost breast plans, was wrongly calculated. This bug was solved in the third year. The number of positioning problems almost stayed the same in the third year, so the improvement in patient positioning is likely to be caused by the use of extra imaging. FM attributed to positioning problems in H&N patients decreased from 9.1 % in the first year to 3.9 % in the second year (highlighted in green in Table 2). This number is gradually getting worse in the third and the fourth year. We are currently looking into this thoroughly to see if retraining the RTT’s is needed or if there could be other reasons, for instance small changes in immobilization strategies. After the introduction of immobilization with calculated couch parameters and an SGRT solution in the third year, FM attributed to patient positioning decreased further for most pathologies, with the largest effect seen for extremities where FM due to patient positioning decreased from 11.2 % in the third year to 4.3 % in the fourth year (highlighted in pink in Table 2).
      Analysis of the results after the first year also showed many failing measurements due to anatomic changes for prostate, rectum, upper abdomen and esophagus patients, mostly weight loss and change in intestinal contents. Therefore, the threshold for follow-up and education by a dietician has been lowered. In Belgium, as part of the cancer plan, funding for dieticians by hospitals with an oncology (primary) care program has been granted.
      FM related to anatomic changes in rectum and prostate patients have been reduced from 10.2 % and 6.7 % in the first year to 4.0 % and 3.3 % in the second year respectively (highlighted in blue in Table 2). The small number of positioning problems in these patients is probably due to the lower sensitivity of transit in vivo dosimetry for these treatment sites [
      • Bojechko C.
      • Ford E.C.
      Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations.
      ,
      • Bossuyt E.
      • Weytjens R.
      • Nevens D.
      • De Vos S.
      • Verellen D.
      Evaluation of automated pre-treatment and transit in-vivo dosimetry in radiotherapy using empirically determined parameters.
      ]. Additionally, a retrospective study was conducted to evaluate if additional help with bladder and rectum preparation in prostate cancer patients by home nurses could improve patients’ preparation, but no statistically significant differences could be observed between the test group and the control group receiving the information on bladder and rectum preparation according to the standard protocol [
      • Fiagan Y.
      • Bossuyt E.
      • Nevens D.
      • Dirix P.
      • Theys F.
      • Gevaert T.
      • et al.
      In vivo dosimetry with prostate cancer to assess possible impact of bladder and rectum preparation. 2020; Tech Innov Patient Support.
      ]. Despite the reduction of FM related to anatomic changes in rectum and prostate patients, there was no noticeable reduction in the number of patient plan adjustments, suggesting the impact of our efforts in that area are still quite limited. Also no difference was observed for abdominal and esophageal cancer patients (highlighted in blue in Table 2). For the esophageal patients in the first year more FM were classified with a cause of positioning problems, but it was seen that these positioning problems are often caused by an anatomic change, which is why the classification has changed after the first year. The overall number of FM didn’t have a noticeable change.
      Following the publication of long-term results of the FAST Forward trial [
      • Murray Brunt A.
      • Haviland J.S.
      • Wheatley D.A.
      • Sydenham M.A.
      • Alhasso A.
      • Bloomfield D.J.
      • et al.
      Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial.
      ,
      • Brunt A.M.
      • Haviland J.S.
      • Sydenham M.
      • Agrawal R.K.
      • Algurafi H.
      • Alhasso A.
      • et al.
      Ten-year results of FAST: a randomized controlled trial of 5-fraction whole-breast radiotherapy for early breast cancer.
      ], postoperative ultrahypofractionated breast RT in five fractions with daily online pre-treatment imaging was implemented in the second year replacing a 15-fraction scheme. FM attributed to positioning problems in breast cancer patients decreased from 6.1 % in the second year to 2.8 % in the third year and FM related to anatomical changes (breast swelling) have been reduced from 1.1 % in the second year to 0.3 % in the third year (highlighted in grey in Table 2). There is only a very small difference in patient positioning problems between the first and the second year, so it’s likely to be the introduction of the 5-fraction scheme in the second year that caused the decrease. Also a retrospective study has been conducted [
      • Fiagan Y.
      • Bossuyt E.
      • Machiels M.
      • Nevens D.
      • Billiet C.
      • Poortmans P.
      • et al.
      Comparing treatment uncertainty for ultra- vs. standard-hypofractionated breast radiation therapy based on in-vivo dosimetry.
      ] comparing 2 groups of 203 breast cancer patients, one group treated with 5 fractions with daily pre-treatment imaging in 2020, the other group treated in 2019 with 15 fractions. To investigate the influence of the daily imaging, a subgroup was created with the results of the fractions of 15fx-group, after pre-treatment imaging. A large difference in patient positioning problems could be observed between the 5-fraction group and the 15-fraction group. Comparison with the subgroup receiving online imaging showed almost no difference though, indicating the daily online IGRT correction protocol is the main cause of the difference between the 2 groups. Also, the difference in breast swelling as a cause disappeared, indicating that using pre-treatment imaging leading to more accurate positioning, could also bring swollen breasts back within constraints [
      • Fiagan Y.
      • Bossuyt E.
      • Machiels M.
      • Nevens D.
      • Billiet C.
      • Poortmans P.
      • et al.
      Comparing treatment uncertainty for ultra- vs. standard-hypofractionated breast radiation therapy based on in-vivo dosimetry.
      ].
      Limitations of this study are that the data were interpreted retrospectively and that the data are not suitable for further statistical analysis and are therefore hypothesis-generating. The number of measurements and detected problems varied between pathologies, causing a larger standard deviation and confidence interval for e.g. extremities that had a lower number of measurements. Over the years the number of measurements were roughly the same though. To our knowledge, no literature is available on the expected evolution of the number and causes of FM. Given this, we were unable to determine in advance which deviations and improvements would be clinically relevant. Therefore, we described changes in numbers of FM and interpret them in light of changes in clinical workflow. Some of these changes were based on the interpretation of past measurements (e.g., assistance of dieticians, training of RTT’s) and others on new scientific data (e.g., publication in implementation of Fast Forward scheme). Each interpretation and decision was made after consultation within a group of physicists, physicians, RTT’s and quality managers with expertise. Other reasons for the trends in the results than the ones reported here cannot be excluded. We also did not use SPC techniques to assess results or establish tolerances. Small dosimetric differences due to algorithm problems and systematic deviations cannot be ruled out. Using these methods the sensitivity and specificity of the IVD could maybe be improved. Vendors could facilitate continuous quality improvement using IVD like the ones described in this study and in other studies such as those of Mans et al. [
      • Mans A.
      • Rozendaal R.
      • Janssen T.
      • Damen E.
      • Kaas J.
      • van Mourik A.
      • et al.
      Reduction of systematic dosimetric uncertainties in volumetric modulated arc therapy triggered by patient-specific quality assurance.
      ], Fuangrod et al. [
      • Fuangrod T.
      • Greer P.B.
      • Simpson J.
      • Zwan B.J.
      • Middleton R.H.
      A method for evaluating treatment quality using in vivo EPID dosimetry and statistical process control in radiation therapy.
      ] and Esposito et al. [
      • Esposito M.
      • Ghirelli A.
      • Pini S.
      • Alpi P.
      • Barca R.
      • Fondelli S.
      • et al.
      Clinical implementation of 3D in vivo dosimetry for abdominal and pelvic stereotactic treatments.
      ] by providing better access to dashboard possibilities, pre-configured classification options and automatic analysis tools for large-scale data including SPC and Receiver Operating Curves.
      Despite the limitations of the study, our results suggest that a broad analysis of IVD results cannot only indicate possible items for improvement but also assess and evaluate the impact of adaptations to the clinical workflow and can hence be a powerful tool in the continuous quality improvement of treatments.

      Declaration of Competing Interest

      The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

      References

        • Derreumaux S.
        • Etard C.
        • Huet C.
        • Trompier F.
        • Clairand I.
        • Bottollier-Depois J.-F.
        • et al.
        Lessons from recent accidents in radiation therapy in France.
        Radiat Prot Dosimetry. 2008; 131: 130-135https://doi.org/10.1093/rpd/ncn235
        • Ortiz López P.
        • Cosset J.M.
        • Dunscombe P.
        • Holmberg O.
        • Rosenwald J.C.
        • Pinillos Ashton L.
        • et al.
        ICRP publication 112. A report of preventing accidental exposures from new external beam radiation therapy technologies.
        Ann ICRP. 2009; 39: 1-86
      1. International Atomic Energy Agency. Development of procedures for in vivo dosimetry in radiotherapy. Human Health Report No. 8, IAEA, Vienna, 2013. https://www-pub.iaea.org/MTCD/Publications/PDF/Pub1606_web.pdf.

      2. International Atomic Energy Agency. Accuracy Requirements and Uncertainties in Radiotherapy, Human Health Series No. 31, IAEA, Vienna, 2016. https://www-pub.iaea.org/MTCD/Publications/PDF/P1679_HH31_web.pdf.

        • Noel A.
        • Aletti P.
        • Bey P.
        • Malissard L.
        Detection of errors in individual patients in radiotherapy by systematic in vivo dosimetry.
        Radiother Oncol. 1995; 34: 144-151https://doi.org/10.1016/0167-8140(94)01503-U
        • Mans A.
        • Wendling M.
        • McDermott L.N.
        • Sonke J.J.
        • Tielenburg R.
        • Vijlbrief R.
        • et al.
        Catching errors with in vivo EPID dosimetry.
        Med Phys. 2010; 37: 2638-2644https://doi.org/10.1118/1.3397807
        • Mijnheer B.
        • González P.
        • Olaciregui-Ruiz I.
        • Rozendaal R.
        • van Herk M.
        • Mans A.
        Overview of 3-year experience with large-scale electronic portal imaging device-based 3-dimensional transit dosimetry Pract.
        Radiat Oncol. 2015; 5: e679-e687https://doi.org/10.1016/j.prro.2015.07.001
        • Mijnheer B.
        • Beddar S.
        • Izewska J.
        • Reft C.
        In vivo dosimetry in external beam radiotherapy.
        Med Phys. 2013; 40070903https://doi.org/10.1118/1.4811216
        • Mijnheer B.
        • Jomehzadeh A.
        • González P.
        • Olaciregui-Ruiz I.
        • Rozendaal R.
        • Shokrani P.
        • et al.
        Error detection during VMAT delivery using EPID-based 3D transit dosimetry.
        Phys Med. 2018; 54: 137-145https://doi.org/10.1016/j.ejmp.2018.10.005
        • Mijnheer B.
        • Olaciregui-Ruiz I.
        • Rozendaal R.
        • Spreeuw H.
        • van Herk M.
        • Mans A.
        Current status of 3D EPID-based in vivo dosimetry in The Netherlands Cancer Institute.
        J Phys Conf Ser. 2015; 573https://doi.org/10.1088/1742-6596/573/1/012014
        • Wolfs C.J.A.
        • Brás M.G.
        • Schyns L.E.J.R.
        • Nijsten S.M.J.J.G.
        • van Elmpt W.
        • Scheib S.G.
        • et al.
        Detection of anatomical changes in lung cancer patients with 2D time-integrated, 2D time-resolved and 3D time-integrated portal dosimetry: a simulation study.
        Phys Med Biol. 2017; 62: 6044-6061https://doi.org/10.1088/1361-6560/aa7730
        • Bojechko C.
        • Ford E.C.
        Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations.
        Med Phys. 2015; 42: 6912-6918https://doi.org/10.1118/1.4935093
        • Bawazeer O.
        • Sarasanandarajah S.
        • Herath S.
        • Kron T.
        • Deb P.
        Sensitivity of electronic portal imaging device (EPID) based transit dosimetry to detect inter-fraction patient variations.
        IFMBE Proc. 2018; 68: 477-480https://doi.org/10.1007/978-981-10-9023-3_86
        • Bojechko C.
        • Phillps M.
        • Kalt A.
        • Ford E.C.
        A quantification of the effectiveness of EPID dosimetry and software-based plan verification systems in detecting incidents in radiotherapy.
        Med Phys. 2015; 42: 5363-5369https://doi.org/10.1118/1.4928601
        • Zhuang A.
        • Olch A.
        Sensitivity study of an automated system for daily patient QA using EPID exit dose images.
        J Appl Clin Med Phys. 2018; 19: 114-124https://doi.org/10.1002/acm2.12303
        • Moustakis C.
        • Ebrahimi Tazehmahalleh F.
        • Elsayad K.
        • Fezeu F.
        • Scobioala S.
        A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetry: a step to QA with in vivo EPID dosimetry.
        Strahlenther Onkol. 2020; 196: 182-192https://doi.org/10.1007/s00066-019-01549-z
        • Doolan P.
        • Nikolaou M.
        • Ferentinos K.
        • Anagnostopoulos G.
        Assessment of a commercial EPID dosimetry system to detect radiotherapy treatment errors.
        Biomed Phys Eng Express. 2021; 7047001https://doi.org/10.1088/2057-1976/ac02a5
        • Van Elmpt W.
        • McDermott L.
        • Nijsten S.
        • Wendling M.
        • Lambin P.
        • Mijnheer B.
        A literature review of electronic portal imaging for radiotherapy dosimetry.
        Radiother Oncol. 2008; 88: 289-309https://doi.org/10.1016/j.radonc.2008.07.008
      3. McCurdy B, Greer P, Bedford J. Electronic portal imaging device dosimetry. In: Mijnheer B. Clinical 3D dosimetry in modern radiation therapy. 2017; 169–98. Electronic Portal Imaging Device Dosimetry | 7 | Clinical 3D Dosimetry (taylorfrancis.com).

        • Celi S.
        • Costa E.
        • Wessels C.
        • Mazal A.
        • Fourquet A.
        • Francois P.
        EPID based in vivo dosimetry system: clinical experience and results.
        J Appl Clin Med Phys. 2016; 17: 262-276https://doi.org/10.1120/jacmp.v17i3.6070
        • Mans A.
        • Rozendaal R.
        • Janssen T.
        • Damen E.
        • Kaas J.
        • van Mourik A.
        • et al.
        Reduction of systematic dosimetric uncertainties in volumetric modulated arc therapy triggered by patient-specific quality assurance.
        Phys Imaging Radiat Oncol. 2022; 21: 6-10https://doi.org/10.1016/j.phro.2022.01.001
        • Fuangrod T.
        • Greer P.B.
        • Simpson J.
        • Zwan B.J.
        • Middleton R.H.
        A method for evaluating treatment quality using in vivo EPID dosimetry and statistical process control in radiation therapy.
        Int J Health Care Qual Assur. 2017; 30: 90-102https://doi.org/10.1108/IJHCQA-03-2016-0028
        • Esposito M.
        • Ghirelli A.
        • Pini S.
        • Alpi P.
        • Barca R.
        • Fondelli S.
        • et al.
        Clinical implementation of 3D in vivo dosimetry for abdominal and pelvic stereotactic treatments.
        Radiother and Oncol. 2021; 154: 14-20https://doi.org/10.1016/j.radonc.2020.09.011
        • Bossuyt E.
        • Weytjens R.
        • Nevens D.
        • De Vos S.
        • Verellen D.
        Evaluation of automated pre-treatment and transit in-vivo dosimetry in radiotherapy using empirically determined parameters.
        Phys Imaging Radiat Oncol. 2020; 16: 113-129https://doi.org/10.1016/j.phro.2020.09.011
        • Olch A.J.
        • O’Meara K.
        • Wong K.K.
        First report of the clinical use of a commercial automated system for daily patient QA using EPID exit images.
        Adv Radiat Oncol. 2019; 4: 722-728https://doi.org/10.1016/j.adro.2019.04.001
        • Esposito M.
        • Piermattei A.
        • Bresciani S.
        • Orlandini L.
        • Falco M.
        • Giancaterino S.
        • et al.
        Improving dose delivery accuracy with EPID in vivo dosimetry: results from a multicenter study.
        Strahlenther Onkol. 2021; 197: 633-643https://doi.org/10.1007/s00066-021-01749-6
        • Nailon W.H.
        • Welsh D.
        • McDonald K.
        • Burns D.
        • Forsyth J.
        • Cooke G.
        • et al.
        EPID-based in vivo dosimetry using Dosimetry Check™: overview and clinical experience in a 5-yr study including breast, lung, prostate, and head and neck cancer patients.
        J Appl Clin Med Phys. 2019; 20: 6-16https://doi.org/10.1002/acm2.12441
        • Fiagan Y.
        • Bossuyt E.
        • Nevens D.
        • Dirix P.
        • Theys F.
        • Gevaert T.
        • et al.
        In vivo dosimetry with prostate cancer to assess possible impact of bladder and rectum preparation. 2020; Tech Innov Patient Support.
        Radiat Oncol. 2020; 16https://doi.org/10.1016/j.tipsro.2020.10.005
        • Murray Brunt A.
        • Haviland J.S.
        • Wheatley D.A.
        • Sydenham M.A.
        • Alhasso A.
        • Bloomfield D.J.
        • et al.
        Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial.
        Lancet. 2020; 395: 1613-1626https://doi.org/10.1016/S0140-6736(20)30932-6
        • Brunt A.M.
        • Haviland J.S.
        • Sydenham M.
        • Agrawal R.K.
        • Algurafi H.
        • Alhasso A.
        • et al.
        Ten-year results of FAST: a randomized controlled trial of 5-fraction whole-breast radiotherapy for early breast cancer.
        J Clin Oncol. 2020; 38: 3261-3272https://doi.org/10.1200/jco.19.02750
        • Fiagan Y.
        • Bossuyt E.
        • Machiels M.
        • Nevens D.
        • Billiet C.
        • Poortmans P.
        • et al.
        Comparing treatment uncertainty for ultra- vs. standard-hypofractionated breast radiation therapy based on in-vivo dosimetry.
        Phys Imag Radiat Oncol. 2022; 22: 85-90https://doi.org/10.1016/j.phro.2022.05.003