Abstract
Background and purpose
A sub-mm resolution Complementary Metal Oxide Semiconductor sensor has been developed for stereotactic radiotherapy quality assurance. Herein we evaluate its basic dosimetric performance and its application for linac C-arm stereotactic body radiotherapy (SBRT) plan quality assurance.
Materials and methods
The detector was integrated into its accompanying phantom or in Water Equivalent Plastic (WEP). The measurement reproducibility, stability, dose linearity and dependence on angularity, dose rate and field size were investigated. Clinical plan measurements were compared to our radiotherapy treatment planning system and radiochromic film. Sensitivity to introduced Multi Leaf Collimator (MLC) offsets was evaluated by simulating single MLC offsets in SBRT plans and comparing measurements to expected doses.
Results
Signal reproducibility was within ± 0.1 % and output calibration was stable over a 6 month period. Detector showed good linearity with dose (r2 = 1). Signal decreased by 5 % when dose rate was decreased from 1300 MU/min to 300 MU/min. Output factors agreed within 0.5 % of chamber measurements for 1x1 cm field sizes or greater. Angularity measurements showed good agreement with reference. For measurement of planned clinical doses, gamma pass-rates were 98.5 % ± 2.3 % (treatment planning system reference, 2 %/2mm) and 99.2 % ± 1.0 % (film reference, 2 %,2mm). The detector also showed sensitivity to errors of 1 mm offsets in MLC positioning.
Conclusion
The detector performed well when used for pre-treatment SBRT plan quality assurance, offering a good alternative to radiochromic film.
1. Introduction
Stereotactic Body Radiotherapy (SBRT) treatments are characterised by a high dose per fraction, steep dose gradients and small radiation fields. High dose rate Flattening Filter Free (FFF) radiation beams are often used, and treatments may be delivered using coplanar or non coplanar beam arrangements or Volumetric Modulated Arc Therapy (VMAT). This poses challenges to achieving reliable and sufficiently accurate pre treatment verification of planned radiation doses [
1- Lo S.S.
- Fakiris A.J.
- Chang E.L.
- Mayr N.A.
- Wang J.Z.
- Papiez L.
- et al.
Stereotactic body radiation therapy: a novel treatment modality.
,
2- Timmerman R.D.
- Kavanagh B.D.
- Cho L.C.
- Papiez L.
- Xing L.
Stereotactic body radiation therapy in multiple organ sites.
,
3The physics of small megavoltage photon beam dosimetry.
,
4- Aspradakis M.M.
- Byrne J.P.
- Palmans H.
- Conway J.
- Rosser K.
- Warrington J.A.P.
Small field MV photon dosimetry.
,
5- Lo S.S.
- Foote M.
- Siva S.
- Slotman B.J.
- Teh B.S.
- Guckenberger M.
- et al.
Technical know-how in stereotactic ablative radiotherapy (SABR).
]. Radiochromic film is often considered the detector of choice for SBRT verification, as no volume averaging effects are observed due to its high spatial resolution [
[6]- Fuss M.
- Sturtewagen E.
- De Wagter C.
- Georg D.
Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance.
]. Measurements are also independent of dose rate, energy and angularity [
[7]- Gonzalez-Lopez A.
- Vera-Sanchez J.A.
- Lago-Martin J.D.
Small fields measurements with radiochromic films.
]. However, film exposure and processing can be very labour intensive and require careful calibration, storage and handling [
[8]Radiochromic film dosimetry.
]. A recent report demonstrated favourable accuracy of a set of synthetic diamond detectors for point dose measurements in stereotactic radiation fields [
[9]- Shaw M.
- Lye J.
- Alves A.
- Keehan S.
- Lehmann J.
- Hanlon M.
- et al.
Characterisation of a synthetic diamond detector for end to end dosimetry in stereotactic body radiotherapy and radiosurgery.
]. However, single element sensors provide very little information about the broader dose distribution.
The need to develop 2D and 3D radiation detectors with significantly greater spatial resolution has prompted a move towards solid state and liquid filled detector arrays, which contain smaller sensor elements but are still capable of generating large magnitude signals. This has resulted in a small range of digital commercial devices suited to pre treatment verification of stereotactic radiotherapy treatment plans. Characterisation of these detectors in stereotactic radiation beams has been recently reported [
10- Shukaili K.A.
- Petasecca M.
- Newall M.
- Espinoza A.
- Perevertaylo V.L.
- Corde S.
- et al.
A 2D silicon detector array for quality assurance in small field dosimetry: DUO.
,
11Yasui K, Saito Y, Ogawa S, Hayashi N. Dosimetric characterization of a new two-dimensional diode detector array used for stereotactic radiosurgery quality assurance. Int J Radiat Res. 2021; 19: 281-89. doi: 10.3390/cancers7030844.
,
12- Markovic M.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Papanikolaou N.
Characterization of a two-dimensional liquid-filled ion chamber detector array used for verification of the treatments in radiotherapy.
], and their suitability for Patient Specific Quality Assurance (PSQA) has been demonstrated for Cyberknife® Stereotactic Radiosurgery (SRS) [
[13]- Xu Q.
- Huynh K.
- Nie W.
- Rose M.S.
- Chawla A.K.
- Choe K.S.
- et al.
Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy.
] as well as linac based SRS [
14- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
,
15- McCulloch J.
- Pawlowski J.
- Kirby N.
- Rasmussen K.
- Shi Z.
- Myers P.
- et al.
Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.
] and SBRT [
16- Rose M.S.
- Tirpak L.
- Van Casteren K.
- Zack J.
- Simon T.
- Schoenfeld A.
- et al.
Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance.
,
17- Markovic M.
- Narayanasamy G.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Saenz D.
- et al.
Clinical evaluation of a two-dimensional liquid-filled ion chamber detector array for verification of high modulation small fields in radiotherapy.
]. These devices have spatial resolutions of ∼ 2.5 mm. Due to the finite detector size, volume averaging and interpolation between sensor elements is required, which adds measurement uncertainty. Recent publications have demonstrated promising results for the use of complementary metal oxide semiconductor (CMOS) technology for radiation dosimetry [
18- Spang F.J.
- Rosenberg I.
- Hedin E.
- Royle G.
Photon small-field measurements with a CMOS active pixel sensor.
,
19- Beck L.
- Velthuis J.J.
- Fletcher S.
- Haynes J.A.
- Page R.F.
Using a TRAPS upstream transmission detector to verify multileaf collimator positions during dynamic radiotherapy delivery.
]. This has led to the commercial clinical release of a detector consisting of 100,000 CMOS elements with an effective spatial resolution of 0.4 mm. This sensor provides continuous dose profiles without the need for significant volume averaging or dose interpolation between the sensor elements. Its application for PSQA has been demonstrated for Cyberknife® SRS [
[20]- Padelli F.
- Aquino D.
- Fariselli L.
- De Martin E.
IBA myQA SRS detector for cyberKnife robotic radiosurgery quality assurance.
]. Our work evaluates the clinical CMOS detector’s performance for C-arm linac SBRT plan quality assurance.
We report the detector’s basic dosimetric properties including signal reproducibility and stability, relationship between signal reported and dose, dose rate, field size (through measurement of output factors) and angularity. To assess clinical dose measurement performance, 22 SBRT plan doses of 6 MV FFF energy were measured and compared to those calculated using our Treatment Planning System (TPS) and measured using radiochromic film. Additionally, SBRT plans containing a single introduced MLC offset were measured and compared to the calculated doses using gamma analyses for a range of clinically relevant gamma parameters to characterise the detector’s sensitivity.
2. Materials & methods
Computed Tomography (CT) scans were acquired using a Philips BigBore CT Scanner. Treatment plans were produced using Raystation version 7.0 or version 10.0 (RaySearch Laboratories, Sweden) using the RaySearch collapsed cone convolution algorithm with a 0.1 cm resolution dose grid. Radiation was generated by 5 matched Elekta Versa HD Linear Accelerators using clinical beams of energy 6 MV FFF (TPR20/10 = 0.679).
The myQA SRS Detector and Phantom are products of IBA Dosimetry (Schwarzenbruck, Germany). Sensor size is 12 × 14 cm, sensor element size is 0.4 × 0.4 mm and spatial resolution is 0.4 mm leaving no spacing between elements. The detector is inserted into a cylindrical phantom with a diameter of 19 cm with a hemispherical end for non-coplanar cranial treatments. The phantom is composed of Acrylonitrile butadiene Styrene (density = 1.06 g/cm3) where the centre of the CMOS detector is positioned at 9.5 cm depth and may be replaced with film or chamber inserts. Film dosimetry was performed using EBT3 film (Ashland Speciality Ingredients). Films were scanned using an Epson X10000 Scanner and analysed using myQA Patients. For ion chamber dosimetry, an NE2571 Farmer chamber (sensitive volume = 0.6 cm3), CC04 chamber (IBA Dosimetry, volume = 0.04 cm3) and a Razor chamber (IBA Dosimetry, volume = 0.01 cm3) were used.
2.1 Basic dosimetric performance
For reproducibility, the % deviation between six successive measurements relative to the first measurement was reported. For stability, output calibration factors acquired each month were compared to previous months. For dose linearity, the CMOS detector and a Farmer chamber were irradiated simultaneously. The CMOS detector was set up with an SSD of 100 cm, measurement depth of 6.7 cm in WEP, and the ion chamber was positioned with its sensitive volume on the central axis downstream of the CMOS detector. The Farmer chamber signal was used to correct detector signal to account for any non linearity in the output of the linear accelerator. For each measurement, the normalised detector signal from a central 4 mm × 4 mm Region of Interest (ROI) was used for our reported values. For dose rate dependence, the detector and a Farmer chamber were set up as for dose linearity measurements. The signal reported from a central 4 mm × 4 mm detector ROI was compared to signal from the Farmer chamber. Signals were normalised to the signal at the maximum nominal dose rate. Dose Per Pulse (DPP) dependence was studied according to the method used by Biasi et al. [
[21]- Biasi G.
- Petrasecca M.
- Guatelli S.
- Hardcastle N.
- Carolan M.
- Perevertaylo V.
- et al.
A novel high-resolution 2D silicon array detector for small field dosimetry with FFF photon beams.
]. Briefly, the detector was set up at 5 cm depth in WEP, and average CMOS detector readings to a central 4 × 4 mm ROI were reported for various SSDs to modulate dose per pulse. These measurements were replicated using a farmer chamber, and a ratio of CMOS reading/chamber reading was used to define the detector’s DPP sensitivity. Measurements for ion recombination effects at each DPP were performed according to the two voltage method [
[22]- Eaton D.J.
- Bass G.
- Booker P.
- Byrne J.
- Duane S.
- Frame J.
- et al.
IPEM code of practice for high energy photon therapy dosimetry based on the NPL absorbed dose calibration service.
]. DPP values were calculated by scaling a literature value for a Elekta versa 6MV FFF [
[23]Xiao Y, Kry SF, Popple R, Yorke E, Papanikolaou N, Stathakis S, Xia P, Huq S, Bayouth J, Galvin J, Yin FF. Flattening filer free accelerators: a report from the AAPM therapy emerging technology assessment work group. J Appl Clin Med Phys. 2015; 16: 12-29 doi: 10.1120%2Fjacmp.v16i3.5219.
] by ion chamber ratios measured under reference and our conditions. For field size dependence, output factors were measured using the CMOS detector and compared to corresponding measurements made using a Farmer chamber from 10x10cm to 5x5cm, a CC04 chamber from 5x5cm to 2x2cm and using a Razor chamber from 2x2cm to 0.5x0.5 cm. Detectors were positioned at 95 cm SSD and 5 cm depth in WEP under full scatter conditions and 100 MUs were delivered. Three sets of measurements were performed for each field size. For field sizes smaller than 3x3cm, beam profiles were acquired using the Razor chamber with a Blue Phantom Smartscan Watertank (IBA Dosimetry) to position the chamber at the centre of the profile. For angular dependence, the CMOS detector was integrated into the SRS Phantom, positioned at machine isocentre and exposed to 100 Monitor Units (MU), 6 MV FFF, 4 cm × 4 cm beam. The average signal to a central 4 mm × 4 mm ROI was recorded for each measurement and was normalised to the signal measured for the exposure at 0° gantry angle. These data were compared to the angular correction data provided by IBA.
2.2 Clinical dose measurement
A total of 22 coplanar VMAT SBRT plans for various anatomical sites (lung, spine, prostate, nodes and liver) were recalculated onto the CT volume of the CMOS detector and CMOS phantom using a 1 mm dose grid. An appropriate isocentre for each plan was chosen and the plans were then delivered to the CMOS detector within the phantom. Measurements were acquired at a frame rate of 2 frames per second and corrected for uniformity, output and angularity. A gantry angle sensor was used to report the incident beam angle required for the angular correction of each frame whilst the detector remained stationary. For analysis, the align data tool within the myQA software was used to remove any additional translational offsets between the planned/film and CMOS doses. Subsequently, a global gamma analysis was performed. For plan measurements using film, the film insert was inserted into the phantom (replacing the detector) prior to the exposure.
2.3 Sensitivity to MLC offsets
The sensitivity of the detector was investigated by studying its ability to detect dose differences corresponding to incorrect MLC configurations within SBRT treatment plans. For four SBRT treatment plans, a single MLC leaf within the target was withdrawn by 1, 2 or 5 mm to simulate the mis calibration of a crucial MLC. Doses from unmodified plans and plans with an introduced offset were calculated using the TPS and measured using the CMOS detector integrated into the phantom. Doses calculated for plans with introduced offsets were compared with doses calculated for unmodified plans to produce a ‘calculated gamma map’ for each introduced offset. This provided the TPS’s estimation of the effect of the offset MLC on dose delivered. Similarly, doses measured for plans with introduced offsets were compared with doses measured for unmodified plans to produce a ‘measured gamma map’ for each offset. This provided the detector’s estimation of the effect of the MLC offset on dose delivered. Doses measured using the CMOS detector for all plans were compared to TPS doses for the unmodified plans using a range of clinically relevant gamma parameters. To establish the clinical effect of these offsets, plan Conformity Indices (CI) were calculated according to the UK SABR Consortium guidelines [
[24]UK SABR consortium. Stereotactic Ablative Body Radiation Therapy (SABR): A Resource (Version 6.1). Brentford, UK: SABR consortium; 2019.
].
4. Discussion
The CMOS detector’s dose linearity is comparable to that reported for other devices, and results were acquired through similar methodologies (although not all studies used FFF beams) [
11Yasui K, Saito Y, Ogawa S, Hayashi N. Dosimetric characterization of a new two-dimensional diode detector array used for stereotactic radiosurgery quality assurance. Int J Radiat Res. 2021; 19: 281-89. doi: 10.3390/cancers7030844.
,
12- Markovic M.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Papanikolaou N.
Characterization of a two-dimensional liquid-filled ion chamber detector array used for verification of the treatments in radiotherapy.
,
20- Padelli F.
- Aquino D.
- Fariselli L.
- De Martin E.
IBA myQA SRS detector for cyberKnife robotic radiosurgery quality assurance.
]. Signal reproducibility was similar to that reported elsewhere for Cyberknife® beams [
[20]- Padelli F.
- Aquino D.
- Fariselli L.
- De Martin E.
IBA myQA SRS detector for cyberKnife robotic radiosurgery quality assurance.
]. Dose rate dependence was consistent with other reports of the same device [
[20]- Padelli F.
- Aquino D.
- Fariselli L.
- De Martin E.
IBA myQA SRS detector for cyberKnife robotic radiosurgery quality assurance.
], and greater than that reported for the SRS MapCheck, for which deviations of < 0.5 % compared to a chamber have been reported for a similar range of dose rates [
11Yasui K, Saito Y, Ogawa S, Hayashi N. Dosimetric characterization of a new two-dimensional diode detector array used for stereotactic radiosurgery quality assurance. Int J Radiat Res. 2021; 19: 281-89. doi: 10.3390/cancers7030844.
,
14- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
]. Our linacs delivered almost all the MUs of our SBRT plans at dose rates exceeding 1000 MU/min, reducing the significance of this dependency. CMOS signal decreased with decreasing DPP, a behaviour observed for detectors of similar composition [
[25]Effects of radiation damage on p-type silicon detectors.
]. This sensitivity variation was ± 2.5 % across the range we studied, which is similar to comparable devices for 6 MV FFF [
[21]- Biasi G.
- Petrasecca M.
- Guatelli S.
- Hardcastle N.
- Carolan M.
- Perevertaylo V.
- et al.
A novel high-resolution 2D silicon array detector for small field dosimetry with FFF photon beams.
]. Due to the geometry of the phantom and isocentric VMAT nature of the plans, DPP effects would generally not be significant but should be considered when interpreting measurements with large offsets. The CMOS detector showed a similar field size dependence to the SRS Mapcheck [
[14]- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
] and the PTW Octavius [
[12]- Markovic M.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Papanikolaou N.
Characterization of a two-dimensional liquid-filled ion chamber detector array used for verification of the treatments in radiotherapy.
] for a similar range of field sizes. Typically, these devices measure output factors within 1 % of ion chamber measurements down to field sizes of 1 cm × 1 cm or 0.5 cm × 0.5 cm, at which deviations of 2–3 % are observed [
10- Shukaili K.A.
- Petasecca M.
- Newall M.
- Espinoza A.
- Perevertaylo V.L.
- Corde S.
- et al.
A 2D silicon detector array for quality assurance in small field dosimetry: DUO.
,
12- Markovic M.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Papanikolaou N.
Characterization of a two-dimensional liquid-filled ion chamber detector array used for verification of the treatments in radiotherapy.
,
14- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
]. The CMOS detector showed an angular dependence like that observed for the Octavius 1000 SRS [
[12]- Markovic M.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Papanikolaou N.
Characterization of a two-dimensional liquid-filled ion chamber detector array used for verification of the treatments in radiotherapy.
] and the SRS Mapcheck [
[14]- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
] for equivalent energies and similar field sizes. All these devices have methods of correcting for this non isotropy, such as gantry mounted inclinometers, which either apply a direct correction to measured signal or facilitate the rotation of the detector such that it is always perpendicular to the gantry. We concluded the CMOS detector’s gantry angle sensor provided sufficient angular correction to measured doses.
The detector demonstrated very good agreement with film and TPS for the measurement of SBRT doses, indicating the detector’s dose rate and DPP dependence weren’t significant. Literature examples of similar detectors used for stereotactic radiotherapy PSQA employed a variety of delivery techniques such as static IMRT fields [
[17]- Markovic M.
- Narayanasamy G.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Saenz D.
- et al.
Clinical evaluation of a two-dimensional liquid-filled ion chamber detector array for verification of high modulation small fields in radiotherapy.
], non coplanar arcs [
[15]- McCulloch J.
- Pawlowski J.
- Kirby N.
- Rasmussen K.
- Shi Z.
- Myers P.
- et al.
Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.
], mixture of coplanar and non coplanar arcs [
14- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
,
16- Rose M.S.
- Tirpak L.
- Van Casteren K.
- Zack J.
- Simon T.
- Schoenfeld A.
- et al.
Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance.
] and Cyberknife deliveries [
13- Xu Q.
- Huynh K.
- Nie W.
- Rose M.S.
- Chawla A.K.
- Choe K.S.
- et al.
Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy.
,
20- Padelli F.
- Aquino D.
- Fariselli L.
- De Martin E.
IBA myQA SRS detector for cyberKnife robotic radiosurgery quality assurance.
]. Furthermore, several treatment sites were studied including brain lesions [
13- Xu Q.
- Huynh K.
- Nie W.
- Rose M.S.
- Chawla A.K.
- Choe K.S.
- et al.
Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy.
,
14- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
,
15- McCulloch J.
- Pawlowski J.
- Kirby N.
- Rasmussen K.
- Shi Z.
- Myers P.
- et al.
Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.
,
20- Padelli F.
- Aquino D.
- Fariselli L.
- De Martin E.
IBA myQA SRS detector for cyberKnife robotic radiosurgery quality assurance.
] and a range of SBRT sites [
16- Rose M.S.
- Tirpak L.
- Van Casteren K.
- Zack J.
- Simon T.
- Schoenfeld A.
- et al.
Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance.
,
17- Markovic M.
- Narayanasamy G.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Saenz D.
- et al.
Clinical evaluation of a two-dimensional liquid-filled ion chamber detector array for verification of high modulation small fields in radiotherapy.
]. These studies used tight gamma parameters to analyse their dose planes due to the greater importance of dosimetric accuracy for stereotactic treatments. We favoured the use of 2 %/2mm due to the high agreement between the CMOS detector, film & planning system as well as the small variation in pass rates. Although these parameters were also favoured in other articles [
[17]- Markovic M.
- Narayanasamy G.
- Stathakis S.
- Mavroidis P.
- Jurkovic I.A.
- Saenz D.
- et al.
Clinical evaluation of a two-dimensional liquid-filled ion chamber detector array for verification of high modulation small fields in radiotherapy.
], our results demonstrated that the use of 3 %/1mm would also be suitable for this detector, as was used in other publications [
13- Xu Q.
- Huynh K.
- Nie W.
- Rose M.S.
- Chawla A.K.
- Choe K.S.
- et al.
Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy.
,
14- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
,
15- McCulloch J.
- Pawlowski J.
- Kirby N.
- Rasmussen K.
- Shi Z.
- Myers P.
- et al.
Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.
,
16- Rose M.S.
- Tirpak L.
- Van Casteren K.
- Zack J.
- Simon T.
- Schoenfeld A.
- et al.
Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance.
]. Literature pass rates for VMAT QA delivered by a C-arm linac using these devices were 98.9 % using the MapCheck (3 %/1mm, n = 6) [
[14]- Ahmed S.
- Zhang G.
- Moros E.G.
- Feygelman V.
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
], 96.0 % also for the MapCheck (3 %/1mm, n = 32) [
[16]- Rose M.S.
- Tirpak L.
- Van Casteren K.
- Zack J.
- Simon T.
- Schoenfeld A.
- et al.
Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance.
] and 99.3 % for the Octavius (3 %/1mm, n = 15) [
[15]- McCulloch J.
- Pawlowski J.
- Kirby N.
- Rasmussen K.
- Shi Z.
- Myers P.
- et al.
Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.
]. These are comparable to the CMOS detector pass rates and suggests that sample size as well as range of treatment sites and delivery styles are an important consideration when attempting to interpret these statistics. Our measured pass rates were significantly greater than the rates reported for the same detector when used to measure significantly non coplanar Cyberknife® deliveries [
[20]- Padelli F.
- Aquino D.
- Fariselli L.
- De Martin E.
IBA myQA SRS detector for cyberKnife robotic radiosurgery quality assurance.
].
The use of gamma parameters 2 %/2mm offered good sensitivity to MLC offsets with minimal variation. The detector’s resolution provides clear visualisation of the gamma map manifestation of the MLC errors. This is a very useful visual aid to assist the operator in diagnosing a mis calibrated MLC, which reenforces the benefits of careful inspection of pairs of dose distributions and their corresponding gamma maps to assist the investigation of suspicious results. Similar work has been published for the Octavius 1000 SRS [
[26]- O’Connor P.
- Seshadri V.
- Charles P.
Detecting MLC errors in stereotactic radiotherapy plans with a liquid filled ionization chamber array.
]. In this study, MLC offsets of up to 0.5 mm were introduced into intracranial SRS plans. These offsets were applied to half the MLCs on the X1 and X2 banks. These MLC errors resulted in a reduction in gamma pass rates, but in terms of clinical significance, the effect on Mean PTV dose was small (<2%) and the effect on CI was inconclusive. Our observed increases in Mean PTV dose difference and CI suggest single MLC offsets of the magnitudes studied could have a significant clinical impact. The UK SABR Consortium recommends CIs for SBRT plans be within the range 1.1 – 1.25, depending on target volume size [
[24]UK SABR consortium. Stereotactic Ablative Body Radiation Therapy (SABR): A Resource (Version 6.1). Brentford, UK: SABR consortium; 2019.
]. Clearly however, the detector’s ability to detect MLC offsets having clinical significance depends on the original plan, it’s quality and department tolerances.
We conclude that the CMOS detector provides a good alternative to film for SBRT QA, having provided consistent clinical dose measurements without requiring inter element dose interpolation or significant volume averaging. The detector also shows promise for being sensitive to MLC offsets that could result in clinically significant dose differences.