1. Introduction
In recent years the number of machines capable of delivering cranial and extra-cranial stereotactic radiotherapy treatments (SBRT) has substantially increased. However, such developments have not been accompanied by dedicated training in terms of small field dosimetry. This may lead to a possible increase of those dosimetric uncertainties that represent the major features of small beams in comparison to wide fields [
1- Das I.J.
- Ding G.X.
- Ahnesjö A.
Small fields: nonequilibrium radiation dosimetry.
,
2- Bouchard H.
- Seuntjens J.
- Duane S.
- Kamio Y.
- Palmans H.
Detector dose response in megavoltage small photon beams. I. Theoretical concepts.
,
3The physics of small megavoltage photon beam dosimetry.
]. Sharing of the measured data in terms of output factors (OF) for small fields, with regard to many different combinations of machines and dosimeters [
4- Russo S.
- Reggiori G.
- Cagni E.
- Clemente S.
- Esposito M.
- Falco M.D.
- et al.
Small field output factors evaluation with a microDiamond detector over 30 Italian centers.
,
5- Mancosu P.
- Pasquino M.
- Reggiori G.
- Masi L.
- Russo S.
- Stasi M.
Dosimetric characterization of small fields using a plastic scintillator detector: a large multi center study.
,
6- Cagni E.
- Russo S.
- Reggiori G.
- Bresciani S.
- Fedele D.
- Iori M.
- et al.
Technical note: multicenter study of TrueBeam FFF beams with a new stereotactic diode: can a common small field signal ratio curve be defined?.
,
7- Masi L.
- Russo S.
- Francescon P.
- Doro R.
- Frassanito M.C.
- Fumagalli M.L.
- et al.
CyberKnife beam output factor measurements: a multi-site and multi-detector study.
,
8- Russo S.
- Masi L.
- Francescon P.
- Frassanito M.C.
- Fumagalli M.L.
- Marinelli M.
- et al.
Multicenter evaluation of a synthetic single-crystal diamond detector for CyberKnife small field size output factors.
], was promoted with a specific working group focused on SBRT. Other had similar experiences [
9Beierholm AR; Behrens CF, Sibolt P, Rønde HS, Biancardo SBN, Aznar MC et al. Reference dosimetry and small-field dosimetry in external beam radiotherapy: Results from a Danish intercomparison study. DTU Nutech 2014 No8(EN); 5–26.
,
10- Followill D.S.
- Kry S.F.
- Qin L.
- Lowenstein J.
- Molineu A.
- Alvarez P.
- et al.
The Radiological Physics Center's standard dataset for small field size output factors.
], all with the aim to quantify systematic inaccuracies of small fields dosimetry.
Small field dosimetry must account for the lack of charged particle equilibrium (CPE) in the lateral direction. Moreover the dimension of the active volume and the material composition and density of the detector itself may influence the response in a different way when compared with non-small radiation fields [
11- Underwood T.S.A.
- Winter H.C.
- Hill M.A.
- Fenwick J.D.
Detector density and small field dosimetry: Integral versus point dose measurement schemes.
,
12- Francescon P.
- Cora S.
- Cavedon C.
Total scatter factors of small beams: a multidetector and Monte Carlo study.
]. In order to solve these issues, the International Atomic Energy Agency (IAEA) and the American Association of Physicists in Medicine (AAPM) proposed a new formalism for small field dosimetry in 2008 [
[13]- Alfonso R.
- Andreo P.
- Capote R.
- Huq M.S.
- Kilby W.
- Kjäll P.
- et al.
A new formalism for reference dosimetry of small and nonstandard fields.
]. In this document, the concept of output correction factor
k for obtaining the dose-to-water ratio between a clinical field (f
clin) and a machine-specific reference field (f
msr), based on the measured detector readings ratio (OF
det) has been introduced. The correction factors can be derived either by a direct calibration of the specific dosimeter in the two fields against a primary standard or against a passive dosimeter such as alanine, Thermoluminescent Dosimeters (TLDs) and radiochromic films, or by a Monte Carlo (MC) simulation [
14- Francescon P.
- Cora S.
- Satariano N.
Calculation of (k(Q(clin), Q(msr))(f(clin), f(msr)) for several small detectors and for two linear accelerators using Monte Carlo simulations.
,
15- García-Garduño O.A.
- Lárraga-Gutiérrez J.M.
- Rodríguez-Villafuerte M.
- Martínez-Dávalos A.
- Celis M.A.
Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom.
]. Moreover, in small beams the positioning accuracy of the collimating devices is also important, since e.g. 1 mm error in a 1 × 1 cm
2 field setting can result in a 2–4% error in the measured dose or OF [
[6]- Cagni E.
- Russo S.
- Reggiori G.
- Bresciani S.
- Fedele D.
- Iori M.
- et al.
Technical note: multicenter study of TrueBeam FFF beams with a new stereotactic diode: can a common small field signal ratio curve be defined?.
].
The main objective of this experience was to assess the consistency in small-field OF determination among a large number of centers, equipped with a wide variety of machines, using different new generation detectors. This multicenter study is a secondary quality assurance (QA) dataset and may be used as a robust baseline for a true audit focused on small-field dosimetry.
2. Methods and materials
A total of thirty centers were enrolled in this study, twenty-four equipped with linacs of different manufactures and models (Varian and Elekta) and six with radiosurgery units, and focused on small-field OF determination using center-specific routine detectors and three new generation detectors: PTW 60019 microDiamond, Exradin W1 Plastic Scintillator (W1 PSD) and IBA new unshielded silicon Razor diode.
2.1 Linac study
Multi-centers studies were performed in four steps:
- 1.
Twelve Varian and twelve Elekta centers performed measurements of the detector readings ratio (OF
det), using f
clin and f
msr, for nominal square field sizes (NFS) ranging from 0.6 cm to 10 cm, employing in-house routinely used detectors and the PTW 60019 microDiamond. A comparison of experimental data obtained with in-house and microDiamond detectors was performed [
[4]- Russo S.
- Reggiori G.
- Cagni E.
- Clemente S.
- Esposito M.
- Falco M.D.
- et al.
Small field output factors evaluation with a microDiamond detector over 30 Italian centers.
];
- 2.
The same centers measured TPR
20/10, in-plane and cross-plane dose profiles of 0.8 × 0.8 cm
2 field and OF
det for NFS ranging from 0.8 cm to 10 cm, using W1 PSD and correcting for the Cerenkov effect as proposed by the manufacturer [
[5]- Mancosu P.
- Pasquino M.
- Reggiori G.
- Masi L.
- Russo S.
- Stasi M.
Dosimetric characterization of small fields using a plastic scintillator detector: a large multi center study.
];
- 3.
Each center was asked to measure in-plane and cross-plane dose profiles and OF
det for square field size ranging from 0.6 cm to 5 cm using Razor diodes [
[16]- Talamonti C.
- Russo S.
- Falco M.D.
- Bartoli A.
- Cagni E.
- Strigari L.
- et al.
Relative Signal Ratios using an unshielded silicon detector: data from 30 centers [abstract].
]. According to Cranmer-Sargison [
[18]- Cranmer-Sargison G.
- Charles P.H.
- Trapp J.V.
- Thwaites D.I.
A methodological approach to reporting corrected small field relative outputs.
], the effective field size (EFS) was calculated as: EFS =
where A and B correspond to the in-plane and cross-plane FWHM values. OF
det, reported as a function of both NFS and EFS, were used to determine a fit of the empirical data, following the equation by Sauer [
[19]Measurement of output ratios for small photon fields.
]:
(1)
where OF is the detector signal ratio; FS is the field size (NFS or EFS); P
∞, S
∞, l, b and n are fit coefficients. In detail, P
∞ represents the maximum primary dose component; S
∞ represents the maximum scatter component. The point (detector reading SR = 1; FS = 30 mm) was considered as a boundary condition.
- 4.
Step 3 was repeated focusing on eight Varian TrueBeam centers [
[6]- Cagni E.
- Russo S.
- Reggiori G.
- Bresciani S.
- Fedele D.
- Iori M.
- et al.
Technical note: multicenter study of TrueBeam FFF beams with a new stereotactic diode: can a common small field signal ratio curve be defined?.
].
The beam energy was 6 MV for steps 1–3, and 10 MV FFF (flattening filter free) for step 4, respectively. The set-up conditions were 10 cm depth in water phantom and SSD = 90 cm. The FS were defined only by the jaws with the MLC fully retracted for the Varian linacs and by both secondary jaws and MLC for the Elekta linacs. The fmrs was 10 × 10 cm2 in the first two steps and 3 × 3 cm2 in the third one. The measurements were performed using two microDiamonds and two Razor diodes in order to speed up the process. The National Institute of Ionizing Radiation Metrology (INMRI) of the Italian institute for new technologies, energy and the environment (ENEA) carried out a complete characterization of both diamond dosimeters and Razor diodes to ensure the dosimetric equivalence of the detectors. Only one W1 detector was used.
Alfonso [
[13]- Alfonso R.
- Andreo P.
- Capote R.
- Huq M.S.
- Kilby W.
- Kjäll P.
- et al.
A new formalism for reference dosimetry of small and nonstandard fields.
] correction factors
were not considered for any of the studied detectors. Details on Linac models involved in the studies have been previously specified [
4- Russo S.
- Reggiori G.
- Cagni E.
- Clemente S.
- Esposito M.
- Falco M.D.
- et al.
Small field output factors evaluation with a microDiamond detector over 30 Italian centers.
,
5- Mancosu P.
- Pasquino M.
- Reggiori G.
- Masi L.
- Russo S.
- Stasi M.
Dosimetric characterization of small fields using a plastic scintillator detector: a large multi center study.
,
6- Cagni E.
- Russo S.
- Reggiori G.
- Bresciani S.
- Fedele D.
- Iori M.
- et al.
Technical note: multicenter study of TrueBeam FFF beams with a new stereotactic diode: can a common small field signal ratio curve be defined?.
].
2.2 Radiosurgery units study
Eight radiosurgery units including all clinically available models were enrolled [
[7]- Masi L.
- Russo S.
- Francescon P.
- Doro R.
- Frassanito M.C.
- Fumagalli M.L.
- et al.
CyberKnife beam output factor measurements: a multi-site and multi-detector study.
]. Inter-linac beam quality (TPR
20/10) variations were below 2%. For each unit, OF
det were measured for fixed cone diameters from 5 to 60 mm using PTW 60017 unshielded silicon diode (routine-used detector for all centers), PTW 60019 microDiamond, W1 PSD and IBA Razor diode. Measurements protocol has been previously described [
[9]Beierholm AR; Behrens CF, Sibolt P, Rønde HS, Biancardo SBN, Aznar MC et al. Reference dosimetry and small-field dosimetry in external beam radiotherapy: Results from a Danish intercomparison study. DTU Nutech 2014 No8(EN); 5–26.
]. Published MC output correction factors specific for radiosurgery unit [
[9]Beierholm AR; Behrens CF, Sibolt P, Rønde HS, Biancardo SBN, Aznar MC et al. Reference dosimetry and small-field dosimetry in external beam radiotherapy: Results from a Danish intercomparison study. DTU Nutech 2014 No8(EN); 5–26.
] were applied to PTW 60017 measured OF
det obtaining the field output factor for each cone size [
[13]- Alfonso R.
- Andreo P.
- Capote R.
- Huq M.S.
- Kilby W.
- Kjäll P.
- et al.
A new formalism for reference dosimetry of small and nonstandard fields.
] as:
(2)
For each radiosurgery unit, microDiamond, W1 PSD and Razor measured OF
det, were compared to PTW 60017 unshielded silicon diode MC corrected values [
[17]- Russo S.
- Masi L.
- Dicarolo P.R.
- Doro R.
- De Martin E.
- Fumagalli M.L.
- et al.
New Razor silicon diode for Cyber Knife small beam relative dosimetry: a multi-site evaluation.
]. The consistency of data over different units was evaluated calculating the ratio K
det for each cone size and detector [
[7]- Masi L.
- Russo S.
- Francescon P.
- Doro R.
- Frassanito M.C.
- Fumagalli M.L.
- et al.
CyberKnife beam output factor measurements: a multi-site and multi-detector study.
] as:
(3)
Det = PTW 60019, W1 PSD, Razor.
K
det is an estimation of the correction factor
needed for the specific detector.
is expected to show small changes (<2%) for the same detector among different radiosurgery units and models [
20- Francescon P.
- Kilby W.
- Satariano N.
- Cora S.
Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife system.
,
21- Francescon P.
- Kilby W.
- Noll J.M.
- Masi L.
- Satariano N.
- Russo S.
Monte Carlo simulated corrections for beam commissioning measurements with circular and MLC shaped fields on the CyberKnife M6 system: a study including diode, microchamber, point scintillator, and synthetic microdiamond detectors.
].
Since our first aim was to evaluate the different detectors performance in a multicentre context, complete results have been collected only for fixed collimators.
For IRIS variable aperture collimator, OF were measured also with PTW 60019 and PTW 60017 [
[8]- Russo S.
- Masi L.
- Francescon P.
- Frassanito M.C.
- Fumagalli M.L.
- Marinelli M.
- et al.
Multicenter evaluation of a synthetic single-crystal diamond detector for CyberKnife small field size output factors.
] but not with W1 PSD. Moreover, these data were not included in the final analysis due to the increased uncertainty in field size diameter defined with the IRIS collimator.
4. Discussion
The lack of a national audit program for small-field dosimetry, performed by independent institutions on a regular basis, boosted the need for a multi-center dosimetric intercomparison to identify systematic dosimetric uncertainties.
In the context of a working group dedicated to small-field dosimetry, the main focus was the OF determination for different types of linac and radiosurgery systems, in a multi-site and multi-detector approach. Output corrections factors, to be applied to experimentally determined ratios of detectors readings, were used only for radiosurgery units diode results, for which well established published MC values were available [
[20]- Francescon P.
- Kilby W.
- Satariano N.
- Cora S.
Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife system.
]. The application of correction factors to linac results was not possible in consideration of the multiple models of linacs and the large variety of routinely-used detectors.
In a preliminary investigation [
[4]- Russo S.
- Reggiori G.
- Cagni E.
- Clemente S.
- Esposito M.
- Falco M.D.
- et al.
Small field output factors evaluation with a microDiamond detector over 30 Italian centers.
], OF measurements with routinely used detectors and nominal FS were re-collected with deviations among centers up to 10% for the 1 × 1 cm
2 field size. The use of microDiamond in the same condition reduced the SD of the OF by about 50%, being below 2% down to 1 × 1 cm
2 FS, in good agreement with the results obtained by on-site quality audits performed by the Radiological Physics Center in the USA [
[10]- Followill D.S.
- Kry S.F.
- Qin L.
- Lowenstein J.
- Molineu A.
- Alvarez P.
- et al.
The Radiological Physics Center's standard dataset for small field size output factors.
]. The use of plastic scintillator detector still yielded a large variability for FS below 2 × 2 cm
2, with limited spread of data (SD <1.5%) for FS greater than 2 × 2 cm
2 [
[5]- Mancosu P.
- Pasquino M.
- Reggiori G.
- Masi L.
- Russo S.
- Stasi M.
Dosimetric characterization of small fields using a plastic scintillator detector: a large multi center study.
]. Using the NFS for data reporting, the comparison between OF measured by different detectors could be affected by differences in the calibration of the collimating devices performed at different times [
[4]- Russo S.
- Reggiori G.
- Cagni E.
- Clemente S.
- Esposito M.
- Falco M.D.
- et al.
Small field output factors evaluation with a microDiamond detector over 30 Italian centers.
].
Regarding the average TPR
20/10 value for 6 MV beam measured with the PSD, the results were consistent with the reference value of 0.676 reported by the IAEA 398 [
[22]Absorbed Dose Determination in External Beam Radiotherapy. An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, Technical Report Series No. 398, IAEA, Vienna, 2000.
].
This exploratory phase resulted in better defined guidelines on how to design our subsequent studies. A multicentric [
[6]- Cagni E.
- Russo S.
- Reggiori G.
- Bresciani S.
- Fedele D.
- Iori M.
- et al.
Technical note: multicenter study of TrueBeam FFF beams with a new stereotactic diode: can a common small field signal ratio curve be defined?.
] study on small-field dosimetry was then performed using the 10 FFF MV beam on 8 TrueBeam linacs, with the awareness that EFS measurements were mandatory when comparing OF data over different centers. A strong relationship between OF
det and the EFS was obtained and its suitability in evaluating uncorrect OF was assessed indicating the Razor as a good detector for small-beam OF evaluation.
Furthermore, semi-empirical output correction factors, K
det, for microDiamond, plastic scintillator and Razor detectors relative to MC corrected silicon diode [
7- Masi L.
- Russo S.
- Francescon P.
- Doro R.
- Frassanito M.C.
- Fumagalli M.L.
- et al.
CyberKnife beam output factor measurements: a multi-site and multi-detector study.
,
8- Russo S.
- Masi L.
- Francescon P.
- Frassanito M.C.
- Fumagalli M.L.
- Marinelli M.
- et al.
Multicenter evaluation of a synthetic single-crystal diamond detector for CyberKnife small field size output factors.
] were evaluated, in the setting of a multicentric study on radiosurgery systems. Radiosurgery K
det values determined by our multisite approach showed high consistency among centers. This uniformity in results emphasizes the value of a multi-center study as a useful tool to identify inaccurate data from a single center. This was probably the case for the single center showing deviating results for 0.5 cm cone size for two of the analysed detectors. The center was contacted to check the data but measurements could not be repeated because in the meantime the system had been uninstalled. Since we were not able to clarify the cause of this outlying result, it was not removed from the final analysis. The semi-empirical output correction factors were confirmed by MC calculated output correction factors within 1% for the radiosurgery unit M6 model, published after the completion of our study [
[21]- Francescon P.
- Kilby W.
- Noll J.M.
- Masi L.
- Satariano N.
- Russo S.
Monte Carlo simulated corrections for beam commissioning measurements with circular and MLC shaped fields on the CyberKnife M6 system: a study including diode, microchamber, point scintillator, and synthetic microdiamond detectors.
]. The agreement with independently calculate MC
is a further validation of the adopted methodology.
In summary, a first step have been completed to develop a national dosimetry audit also in collaboration with ENEA. Our intercomparison studies supported the improvement of accuracy and safety in small-field dosimetry, in particular using extreme caution when selecting the dosimeter and its placement within the field essential to avoid gross errors in the determination of small-field OF data. As soon as, the national dosimetry audit will focus on determining small field OF, any challenges or difficulties in the proposed methodology will be addressed using the experience of this study.