Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy

  • Author Footnotes
    1 Department of Radiation Oncology, Calvary Mater Newcastle, Edith and Platt St, Newcastle, NSW 2300, Australia.
    Joerg Lehmann
    Correspondence
    Corresponding author at: Department of Radiation Oncology, Calvary Mater Newcastle, Edith and Platt St, Newcastle NSW 2300, Australia.
    Footnotes
    1 Department of Radiation Oncology, Calvary Mater Newcastle, Edith and Platt St, Newcastle, NSW 2300, Australia.
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia

    Institute of Medical Physics, School of Physics A28, University of Sydney NSW 2006, Australia

    School of Mathematical and Physical Sciences, The University of Newcastle, University Drive, Callaghan, NSW 2308, Australia

    School of Science, Royal Melbourne Institute of Technology (RMIT) University, 124 La Trobe Street, Melbourne, VIC 3000, Australia
    Search for articles by this author
  • Andrew Alves
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Author Footnotes
    2 Epworth Radiation Oncology, 32 Erin Street, Richmond, VIC 3121, Australia.
    Leon Dunn
    Footnotes
    2 Epworth Radiation Oncology, 32 Erin Street, Richmond, VIC 3121, Australia.
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Maddison Shaw
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia

    School of Science, Royal Melbourne Institute of Technology (RMIT) University, 124 La Trobe Street, Melbourne, VIC 3000, Australia
    Search for articles by this author
  • John Kenny
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Author Footnotes
    3 Department of Radiation Oncology, Peter MacCallum Cancer Center, 305 Grattan Street, Melbourne, VIC 3000, Australia.
    Stephanie Keehan
    Footnotes
    3 Department of Radiation Oncology, Peter MacCallum Cancer Center, 305 Grattan Street, Melbourne, VIC 3000, Australia.
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia

    School of Science, Royal Melbourne Institute of Technology (RMIT) University, 124 La Trobe Street, Melbourne, VIC 3000, Australia
    Search for articles by this author
  • Jeremy Supple
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Author Footnotes
    4 Adem Crosby Centre, Sunshine Coast University Hospital, 6 Doherty Street, Birtinya, QLD 4575, Australia.
    Francis Gibbons
    Footnotes
    4 Adem Crosby Centre, Sunshine Coast University Hospital, 6 Doherty Street, Birtinya, QLD 4575, Australia.
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Sophie Manktelow
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Chris Oliver
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Tomas Kron
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia

    School of Science, Royal Melbourne Institute of Technology (RMIT) University, 124 La Trobe Street, Melbourne, VIC 3000, Australia

    Department of Radiation Oncology, Peter MacCallum Cancer Center, 305 Grattan Street, Melbourne, VIC 3000, Australia
    Search for articles by this author
  • Ivan Williams
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Jessica Lye
    Affiliations
    Australian Clinical Dosimetry Service (ACDS), Australian Radiation Protection and National Safety Agency (ARPANSA), 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
    Search for articles by this author
  • Author Footnotes
    1 Department of Radiation Oncology, Calvary Mater Newcastle, Edith and Platt St, Newcastle, NSW 2300, Australia.
    2 Epworth Radiation Oncology, 32 Erin Street, Richmond, VIC 3121, Australia.
    3 Department of Radiation Oncology, Peter MacCallum Cancer Center, 305 Grattan Street, Melbourne, VIC 3000, Australia.
    4 Adem Crosby Centre, Sunshine Coast University Hospital, 6 Doherty Street, Birtinya, QLD 4575, Australia.
Open AccessPublished:April 20, 2018DOI:https://doi.org/10.1016/j.phro.2018.03.006

      Highlights

      • End-to-end radiotherapy audits have been performed on 94 occasions in Australia.
      • Conformal fields were assessed on an anthropomorphic thorax phantom.
      • Dose after lung measured low for analytical anisotropic algorithm (AAA) calculations.
      • AAA also overestimated dose in tangential fields as used in breast cancer treatments.
      • AAA underestimated dose to points inside lung inhomogeneities.
      • Superposition and convolution algorithms had problems with large angle wedges.

      Abstract

      Background and purpose

      Independent dosimetry audits improve quality and safety of radiation therapy. This work reports on design and findings of a comprehensive 3D conformal radiotherapy (3D-CRT) Level III audit.

      Materials and methods

      The audit was conducted as onsite audit using an anthropomorphic thorax phantom in an end-to-end test by the Australian Clinical Dosimetry Service (ACDS). Absolute dose point measurements were performed with Farmer-type ionization chambers. The audited treatment plans included open and half blocked fields, wedges and lung inhomogeneities. Audit results were determined as Pass Optimal Level (deviations within 3.3%), Pass Action Level (greater than 3.3% but within 5%) and Out of Tolerance (beyond 5%), as well as Reported Not Scored (RNS). The audit has been performed between July 2012 and January 2018 on 94 occasions, covering approximately 90% of all Australian facilities.

      Results

      The audit pass rate was 87% (53% optimal). Fifty recommendations were given, mainly related to planning system commissioning. Dose overestimation behind low density inhomogeneities by the analytical anisotropic algorithm (AAA) was identified across facilities and found to extend to beam setups which resemble a typical breast cancer treatment beam placement. RNS measurements inside lung showed a variation in the opposite direction: AAA under-dosed a target beyond lung and over-dosed the lung upstream and downstream of the target. Results also highlighted shortcomings of some superposition and convolution algorithms in modelling large angle wedges.

      Conclusions

      This audit showed that 3D-CRT dosimetry audits remain relevant and can identify fundamental global and local problems that also affect advanced treatments.

      Keywords

      1. Introduction

      Quality of Radiation Therapy delivery directly impacts the outcome of the treatments delivered to patients. This includes avoiding catastrophic failures but also adherence to the details of dosimetric procedures, which can have a measurable impact on clinical outcomes [
      • Peters L.J.
      • O'Sullivan B.
      • Giralt J.
      • Fitzgerald T.J.
      • Trotti A.
      • Bernier J.
      • et al.
      Critical impact of radiotherapy protocol compliance and quality in the treatment of advanced head and neck cancer: results from TROG 02.02.
      ]. The role of dosimetry audits in the context of clinical trials has been established [
      • Kron T.
      • Haworth A.
      • Williams I.
      Dosimetry for audit and clinical trials: challenges and requirements.
      ,
      • Ibbott G.S.
      • Haworth A.
      • Followill D.S.
      Quality Assurance for Clinical Trials.
      ,
      • Palmer A.
      • Mzenda B.
      • Kearton J.
      • Wills R.
      Analysis of regional radiotherapy dosimetry audit data and recommendations for future audits.
      ,
      • Pettersen M.N.
      • Aird E.
      • Olsen D.R.
      Quality assurance of dosimetry and the impact on sample size in randomized clinical trials.
      ] and dosimetry audits for clinical trials are being performed worldwide [
      • Izewska J.
      • Lechner W.
      • Wesolowska P.
      Global availability of dosimetry audits in radiotherapy: The IAEA dosimetry audit networks database.
      ]. The role of dosimetry audits as a quality assurance tool outside clinical trials is expanding [
      • Clark C.H.
      • Aird E.G.
      • Bolton S.
      • Miles E.A.
      • Nisbet A.
      • Snaith J.A.
      • et al.
      Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
      ,
      • Thomas R.A.S.
      • Bolt M.A.
      • Bass G.
      • Nutbrown R.
      • Chen T.
      • Nisbet A.
      • et al.
      Radiotherapy reference dose audit in the United Kingdom by the National Physical Laboratory: 20 years of consistency and improvements.
      ]. They are part of government efforts to improve and maintain quality in radiation therapy. Participation in dosimetry audits has become a component of licensing processes in some jurisdictions.
      Level III audits commonly use a phantom which is put through the entire chain of procedures that a patient would go through during radiation treatment simulation, planning and delivery [
      • Kron T.
      • Haworth A.
      • Williams I.
      Dosimetry for audit and clinical trials: challenges and requirements.
      ]. Substantial experience in Level III and other comprehensive dosimetry audits has been reported on from the northern hemisphere; including audits in the United Kingdom (UK) [
      • Clark C.H.
      • Aird E.G.
      • Bolton S.
      • Miles E.A.
      • Nisbet A.
      • Snaith J.A.
      • et al.
      Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
      ,
      • Clark C.H.
      • Hussein M.
      • Tsang Y.
      • Thomas R.
      • Wilkinson D.
      • Bass G.
      • et al.
      A multi-institutional dosimetry audit of rotational intensity-modulated radiotherapy.
      ,
      • Hussein M.
      • Tsang Y.
      • Thomas R.A.
      • Gouldstone C.
      • Maughan D.
      • Snaith J.A.
      • et al.
      A methodology for dosimetry audit of rotational radiotherapy using a commercial detector array.
      ,
      • Lee J.
      • Mayles H.
      • Baker C.
      • Jafari S.
      • Distefano G.
      • Clark C.
      UK SABR Consortium lung dosimetry audit; Relative dosimetry results.
      ] and those offered by Imaging and Radiation Oncology Core (IROC) [
      • Followill D.S.
      • Evans D.R.
      • Cherry C.
      • Molineu A.
      • Fisher G.
      • Hanson W.F.
      • et al.
      Design, development, and implementation of the radiological physics center's pelvis and thorax anthropomorphic quality assurance phantoms.
      ,
      • Ibbott G.S.
      • Molineu A.
      • Followill D.S.
      Independent evaluations of IMRT through the use of an anthropomorphic phantom.
      ,
      • Ibbott S.I.
      • Followill D.S.
      The Radiological Physics Center: 40 years of vigilance and quality assurance for NCI sponsored clinical trials.
      ,
      • Kry S.F.
      • Molineu A.
      • Kerns J.R.
      • Faught A.M.
      • Huang J.Y.
      • Pulliam K.B.
      • et al.
      Institutional patient-specific IMRT QA does not predict unacceptable plan delivery.
      ,
      • Pulliam K.B.
      • Followill D.
      • Court L.
      • Dong L.
      • Gillin M.
      • Prado K.
      • et al.
      A six-year review of more than 13,000 patient-specific IMRT QA results from 13 different treatment sites.
      ,
      • Kerns J.R.
      • Followill D.S.
      • Lowenstein J.
      • Molineu A.
      • Alvarez P.
      • Taylor P.A.
      • et al.
      Agreement between institutional measurements and treatment planning system calculations for basic dosimetric parameters as measured by the imaging and radiation oncology core-Houston.
      ,
      • Kry S.F.
      • Dromgoole L.
      • Alvarez P.
      • Leif J.
      • Molineu A.
      • Taylor P.
      • et al.
      Radiation therapy deficiencies identified during on-site dosimetry visits by the imaging and radiation oncology core Houston quality assurance center.
      ]. Audits use different dosimeters, including thermoluminescence detectors (TLD) [
      • Faught A.M.
      • Kry S.F.
      • Luo D.
      • Molineu A.
      • Bellezza D.
      • Gerber R.L.
      • et al.
      Development of a modified head and neck quality assurance phantom for use in stereotactic radiosurgery trials.
      ], optically stimulated luminescence detectors (OSLD) [
      • Kry S.F.
      • Dromgoole L.
      • Alvarez P.
      • Leif J.
      • Molineu A.
      • Taylor P.
      • et al.
      Radiation therapy deficiencies identified during on-site dosimetry visits by the imaging and radiation oncology core Houston quality assurance center.
      ,
      • Lehmann J.
      • Dunn L.
      • Lye J.E.
      • Kenny J.W.
      • Alves A.D.
      • Cole A.
      • et al.
      Angular dependence of the response of the nanoDot OSLD system for measurements at depth in clinical megavoltage beams.
      ,
      • Kerns J.R.
      • Kry S.F.
      • Sahoo N.
      • Followill D.S.
      • Ibbott G.S.
      Angular dependence of the nanoDot OSL dosimeter.
      ], Alanine [
      • Clark C.H.
      • Aird E.G.
      • Bolton S.
      • Miles E.A.
      • Nisbet A.
      • Snaith J.A.
      • et al.
      Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
      ], radiochromic film [
      • Das I.J.
      Radiochromic film: role and applications in radiation dosimetry.
      ], ionization chambers [
      • Hussein M.
      • Tsang Y.
      • Thomas R.A.
      • Gouldstone C.
      • Maughan D.
      • Snaith J.A.
      • et al.
      A methodology for dosimetry audit of rotational radiotherapy using a commercial detector array.
      ,
      • Lye J.
      • Kenny J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      • Alves A.
      • et al.
      A 2D ion chamber array audit of wedged and asymmetric fields in an inhomogeneous lung phantom.
      ,
      • de Prez L.
      • Heukelom S.
      • Jansen B.
      • Jansen W.
      • van de Kamer J.
      • van Klink W.
      • et al.
      An on-site dosimetry audit for high-energy electron beams.
      ] and electronic portal imaging devices [
      • Miri N.
      • Lehmann J.
      • Legge K.
      • Vial P.
      • Greer P.B.
      Virtual EPID standard phantom audit (VESPA) for remote IMRT and VMAT credentialing.
      ,
      • Miri N.
      • Lehmann J.
      • Legge K.
      • Zwan B.J.
      • Vial P.
      • Greer P.B.
      Remote dosimetric auditing for intensity modulated radiotherapy: A pilot study.
      ] the choice of which impacts accuracy and timeframe of result availability. High level audits often rely on volunteer efforts [
      • Clark C.H.
      • Aird E.G.
      • Bolton S.
      • Miles E.A.
      • Nisbet A.
      • Snaith J.A.
      • et al.
      Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
      ], and some comprehensive audits have been cut back due to financial constraints [
      • Kry S.F.
      • Dromgoole L.
      • Alvarez P.
      • Leif J.
      • Molineu A.
      • Taylor P.
      • et al.
      Radiation therapy deficiencies identified during on-site dosimetry visits by the imaging and radiation oncology core Houston quality assurance center.
      ]. Many high level audits nowadays focus on the use of advanced treatment modalities [
      • Ebert M.A.
      • Harrison K.M.
      • Cornes D.
      • Howlett S.J.
      • Joseph D.J.
      • Kron T.
      • et al.
      Comprehensive Australasian multicentre dosimetric intercomparison: issues, logistics and recommendations.
      ,
      • Kron T.
      • Pham D.
      • Roxby P.
      • Rolfo A.
      • Foroudi F.
      Credentialing of radiotherapy centres for a clinical trial of adaptive radiotherapy for bladder cancer (TROG 10.01).
      ,
      • Tsang Y.
      • Carver A.
      • Groom N.
      • Harris C.
      • Faivre-Finn C.
      • Eaton D.
      A multi-centre dosimetry audit on advanced radiotherapy in lung as part of the Isotoxic IMRT study.
      ].
      The presented Level III (end-to-end test) dosimetry audit for 3D conformal radiotherapy (3D-CRT) differed from similar audits in several aspects: Using primary standard calibrated dosimeters and an onsite audit approach with a team exclusively dedicated to radiotherapy audits the audit was able to have lower uncertainties and tighter action levels. Combined with covering almost an entire country and continent, including private and public providers alike and irrespective of clinical trials participation, this meant that the audit did not only give an overview of the status of the dosimetry therein, but it was also able to identify even small problems and trends with certain equipment combinations.

      2. Material and methods

      2.1 Australian Clinical Dosimetry Service

      Starting in 2011 Australia has implemented a national dosimetry audit program with the Australian Clinical Dosimetry Service (ACDS). The program had initially been trialled as a government funded pilot project [
      • Williams I.
      • Kenny J.
      • Lye J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      The Australian Clinical Dosimetry Service: a commentary on the first 18 months.
      ] and is now self-funded (as of January 2017). The ACDS has built their audit system utilizing the work of others [
      • Clark C.H.
      • Aird E.G.
      • Bolton S.
      • Miles E.A.
      • Nisbet A.
      • Snaith J.A.
      • et al.
      Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
      ,
      • Ibbott G.S.
      • Followill D.S.
      • Molineu H.A.
      • Lowenstein J.R.
      • Alvarez P.E.
      • Roll J.E.
      Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials.
      ,
      • IAEA. TECDOC 1583
      Commissioning of radiotherapy treatment planning systems: Testing for typical external beam treatment techniques.
      ] while further improving the methods [
      • Lehmann J.
      • Dunn L.
      • Lye J.E.
      • Kenny J.W.
      • Alves A.D.
      • Cole A.
      • et al.
      Angular dependence of the response of the nanoDot OSLD system for measurements at depth in clinical megavoltage beams.
      ,
      • Williams I.
      • Kenny J.
      • Lye J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      The Australian Clinical Dosimetry Service: a commentary on the first 18 months.
      ,
      • Dunn L.
      • Lye J.
      • Kenny J.
      • Lehmann J.
      • Williams I.
      • Kron T.
      Commissioning of optically stimulated luminescence dosimeters for use in radiotherapy.
      ] and adapting to the dimensions, population densities and technological diversity of Australia and to the rules of the pilot funding [
      • Williams I.
      • Kenny J.
      • Lye J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      The Australian Clinical Dosimetry Service: a commentary on the first 18 months.
      ,
      • Williams I.
      • Kenny J.
      • Lye J.
      • Lehmann J.
      The Australian Clinical Dosimetry Service: the development and delivery of a national dosimetry audit. World congress on medical physics and biomedical engineering May 26–31, 2012.
      ]. The latter included deliverables for a three year time line, a focus on general clinical needs rather than on clinical trials, and inclusion of 3D conformal therapy audits. Details of ACDS Level I and Level II audits have been reported [
      • Lye J.
      • Kenny J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      • Alves A.
      • et al.
      A 2D ion chamber array audit of wedged and asymmetric fields in an inhomogeneous lung phantom.
      ,
      • Lye J.
      • Dunn L.
      • Kenny J.
      • Lehmann J.
      • Kron T.
      • Oliver C.
      • et al.
      Remote auditing of radiotherapy facilities using optically stimulated luminescence dosimeters.
      ]. So have selected findings of the Level II and III audits [
      • Dunn L.
      • Lehmann J.
      • Lye J.
      • Kenny J.
      • Kron T.
      • Alves A.
      • et al.
      National dosimetric audit network finds discrepancies in AAA lung inhomogeneity corrections.
      ]. This work reports comprehensively on the Level III 3D-CRT audit. Following four field trials, February-April 2012, the audit has been performed on 94 occasions between July 2012 and January 2018, covering approximately 90% of all Australian facilities.

      2.2 Onsite audit

      This audit was designed as an onsite audit. Prescribed planning, quality assurance (QA) and delivery were performed by clinical staff from the audited facility while an audit team measured the dose delivered to the phantom. Outreach to radiation therapists through presentations at their national meetings and a publication in their journal helped to closer involve them in the audit process [
      • Lehmann J.
      • Kenny J.
      • Lye J.
      • Kron T.
      • Williams I.
      Radiation therapists and Level III audits by the Australian Clinical Dosimetry Service.
      ].
      The onsite approach was chosen over a postal audit as it allowed the program to start quickly with a single phantom and to use ionization chambers. Being onsite during the audit provided the opportunity for the audit team to observe any problems with the audit procedures. Additionally, an onsite audit team could help with troubleshooting in case of suboptimal audit results, bringing the facility back to high quality patient treatment more quickly.

      2.3 Dosimetry equipment

      Ionisation chambers offer the highest accuracy for field measurements and immediate readout and therefore quick availability of (preliminary) audit result. Farmer type PTW 30013 chambers (Physikalisch Technische Werkstätten, Freiburg, Germany) were selected for this audit, as they were for the ACDS Level Ib audit [
      • Williams I.
      • Kenny J.
      • Lye J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      The Australian Clinical Dosimetry Service: a commentary on the first 18 months.
      ]. This overlap and the chamber’s known quality and robustness were accepted as a trade-off for their larger volume compared to available smaller chambers.
      The “PC electrometer” (Sun Nuclear, Melbourne, FL, USA) was selected for its small size, enabling the audit team to also bring a backup device. The lightweight two channel electrometer was thoroughly tested before deployment. The presence of a second channel allowed for concurrent measurement at two locations in the phantom. Some of the secondary measurement locations were considered interesting but not critical for evaluation and potentially subject to larger uncertainties. For those points the “Reported Not Scored”, RNS category was introduced, which is also used in the ACDS Level II audit [
      • Lye J.
      • Kenny J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      • Alves A.
      • et al.
      A 2D ion chamber array audit of wedged and asymmetric fields in an inhomogeneous lung phantom.
      ]. The electrometer’s log function provided a record of all measurements.
      Facility independent equipment including backup was brought onsite. Ionization chambers and electrometers had been directly calibrated by the Australian Primary Standards Laboratory. See Supplemental Material for details regarding logistics.

      2.4 Audit phantom

      The commercial anthropomorphic “IMRT Thorax Phantom Model 002LFC” (CIRS, Norfolk, VA, USA) had been initially chosen over a custom phantom to reduce the risk of downtime if the phantom was lost or damaged in transport. This choice also allowed facilities to replicate audit conditions when troubleshooting.
      To reduce its weight and size for easier transport, the phantom (Supplementary Fig. 1) was shortened by seven slices. In mid-2016 the commercial phantom was replaced with a custom phantom designed to accommodate both, the here discussed 3D-CRT fields, some of which are continued to be used, and new, intensity modulated and volumetric arc therapy fields for additional checks, which are beyond the scope of this work. The risk of downtime due to loss or damage remains and will be mitigated by maintaining two phantoms.

      2.5 Audit cases and measurement locations

      At the time of the audit set-up (2012) the majority of patients in Australia were treated with 3D-CRT. Conformal treatment plans were chosen to test performance with wedges, asymmetric fields and low density inhomogeneities.
      Measurements were performed as absolute dose measurements using an adapted TRS 398 [
      • IAEA. TRS 398
      Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water.
      ] approach. This included corrections for temperature and pressure. The correction factor kQ was calculated based on the facility provided beam quality information, while standard correction factors ks and kpol were used based on experience with each ionization chamber.
      For efficiency reasons all cases were initially only delivered with standard (flattened) 6 MV beams. Higher energy 3D-CRT beams were introduced in mid-2016 and are not discussed here.
      Case 1 investigated system performance close to reference conditions in a surface distance based setup at 3 cm depth (Point 1). Illustrations of all cases are in Supplementary Figs. 2–5. Table 1 in [
      • Dunn L.
      • Lehmann J.
      • Lye J.
      • Kenny J.
      • Kron T.
      • Alves A.
      • et al.
      National dosimetric audit network finds discrepancies in AAA lung inhomogeneity corrections.
      ] lists plan details. Point 10 was used to assess depth dose accuracy. A measured correction factor of ∼1% was applied to compensate for the increased dose at Point 10 due to the presence of a chamber at Point 1.
      Case 2 tested performance with a wedge on an oblique body surface using a single field plan adapted from [
      • IAEA. TECDOC 1583
      Commissioning of radiotherapy treatment planning systems: Testing for typical external beam treatment techniques.
      ]. Point 1 was the isocenter and prescription point. Point 4 was located upstream and posteriorly, moving it towards the thinner end of the wedge. Point 7 was measured as RNS to gain understanding about the performance of the planning system for points far outside the field.
      Case 3 was a three field plan with an anterior field and two half-beam blocked lateral fields with 30° wedges passing though lung material [
      • IAEA. TECDOC 1583
      Commissioning of radiotherapy treatment planning systems: Testing for typical external beam treatment techniques.
      ]. The main measurement location was Point 5. Additional measurements were taken at Point 8, inside the lung material, and at Point 10, at greater depth for the anterior beam. Each beam was analysed separately.
      Cases 4 and 5 were added to test the modelling of a 60° wedge. The phantom position was not changed between the two cases allowing for assessing the impact of the wedge at a series of points at different depths and at one off-axis point. Case 4 by itself also served in the assessment of the handling of dose behind lung inhomogeneities.

      2.6 Audit scoring

      The variation from the ACDS measured dose was calculated with local reference for each point:
      Dose Variation from ACDSLocal Reference=Facility Stated Dose-ACDS Measured DoseACDS Measured Dose
      (1)


      Audit point results were determined as Pass Optimal Level (deviations within 3.3%), Pass Action Level (greater than 3.3% but within 5%) and Out of Tolerance (beyond 5%), in addition to the above mentioned RNS. The result of the overall audit was equal to that of the worst measurement point.
      Facilities provided their assessment of the current linac output. However, for audit scoring no corrections were made for deviations from 1 cGy/MU. For comparisons across facilities, and as noted with the corresponding figures, the results have often been displayed relative to those of Case 1, Point 1 focussing them on the specific test and eliminating the impact of daily output fluctuations and reference condition accuracy.
      Dose Variationcorrected=Dose Variationuncorrected+1Dose Variationc1p1+1-1
      (2)


      3. Results

      3.1 Linac output/Reference

      The facility reported dose variation from the nominal 1 cGy/MU ranged from −2.0% to 1.3% (Mean ± 1 SD: 0.1 ± 0.6%).
      Planning system accuracy close to reference conditions, measured with Case 1 Point 1, was within “Pass Optimal” for all audits, as shown in Fig. 1.
      Figure thumbnail gr1
      Fig. 1Dose variation at near reference condition (Case 1 Point 1): raw variation as per equation (1) (closed circles, Mean −0.4%, SD 1.2%) and variation corrected by facility reported output (open symbols, Mean −0.5%, SD 1.0%). Here and in all following figures results are broken down by planning and delivery system combination. Eclipse results are further broken down into dose calculation algorithms, as some results differ between them. Results for Pinnacle, Xio and Monaco are not further broken into the encountered respective dose calculation algorithms, as no significant differences have been observed here.

      3.2 PDD modelling

      Fig. 2 shows the variations found for doses measured at larger depths in the phantom, corrected for those found for Point 1 Case 1, which describes the ability of the planning system to correctly model depth dose.
      Figure thumbnail gr2
      Fig. 2PDD investigations: dose variations for points at different depths in the phantom corrected by the dose variation measured in Case 1 Point 1 per Eq. .
      For the larger depth (Case 1 Point 10) the average variation was −0.4 ± 1.3%, indicating that there is no general bias towards either overestimating or underestimating the depth-dose curve. This is supported by the results for the shallower Point 5 of Case 3 (−0.2 ± 0.8%) and as well Case 3 Point 10 (−0.1 ± 2.8%, not shown).

      3.3 Lung inhomogeneity

      The analytical anisotropic algorithm (AAA) overestimated the dose in water equivalent material behind lung (Case 4 Fig. 3), leading to an underdose in the measurement. When investigating multiple points at different depths behind the lung material (Points 2, 3 and 4) there was no depth dependence of the magnitude of the overestimation from the AAA algorithm (1.9 ± 1.0%, 2.2 ± 0.8% and 1.9 ± 0.8%, respectively). However, deviations were larger for the off axis point 5 (3.2 ± 1.2%) which included a longer path through the lung material.
      Figure thumbnail gr3
      Fig. 3Dose variation for points behind lung (Case 4 Points 2–5), corrected by the dose variation measured in Case 1 Point 1 per Eq. .
      Fig. 4 shows exploratory investigations (RNS) into the performance of treatment planning system (TPS) algorithms to accurately predict the dose in lung using a chamber directly inserted into the lung material (Case 3 Point 8). AAA underestimated the dose in this case. For the right lateral (RLAT) field the underestimation was markedly larger than for the left lateral (LLAT) field. Assessment of a point located centrally in the phantom in water equivalent material (Point 5 – Supplemental Fig. 8) showed the overestimation of dose behind the lung for the AAA algorithm as in Case 4 (Fig. 3). The Monaco system also underestimated the dose in lung but equally for both beams, while Xio overestimated the dose in lung for Varian machines for both beams. In both systems dose calculation for the central Point 5 was in agreement with measurement. Small variations seen for Pinnacle/Elekta inside lung material are similar to those in the central point in water equivalent material.
      Figure thumbnail gr4
      Fig. 4Dose variation inside lung: Case 3 Point 8, right lateral (RLAT) and left lateral (LLAT) beams, corrected by the dose variation measured in Case 1 Point 1 per Eq. .
      Results for Point 1 of Case 2, the direct beam path to which passes close by but does not traverse lung tissue, indicated a similar effect as that of lung for the AAA algorithm (Fig. 5). Point 4 of Case 2 caused many out of tolerance results for the audit. Its location close to the field edge required accurate setup and jaw position calibration in addition to correct inhomogeneity handling by the planning system.
      Figure thumbnail gr5
      Fig. 5Dose variations for Case 2. Dose variation at the measurement points has been corrected by the dose variation measured in Case 1 Point 1 per Eq. .

      3.4 Wedges

      Wedged field calculations presented problems for few selected facilities only (Fig. 6). System related trends were seen for the Pinnacle/Elekta combination with a negative bias due to the wedge and for the Xio/Elekta combination with most points grouped around −3 to −4%. A single beam model had results in the opposite direction grouped, suggesting variation in the individual facility implementation of the algorithm.
      Figure thumbnail gr6
      Fig. 6Wedge impact. Difference in dose variations for Cases 4 (no wedge) and 5 (60° wedge).

      3.5 Overall outcomes

      Of the 94 audits 50 (53%) scored “Pass Optimal Level”, 32 (34%) “Pass Action Level” and 12 (13%) “Out Of Tolerance”.
      Recommendations were given regarding the modelling of low-density inhomogeneities (16), photon calibration (6), the handling of wedges in general (5) and off axis (5), depth dose modelling (2), and temperature/pressure correction (2). Additionally, problems occurred when facilities did not follow their internal procedures when preparing or delivering the audit cases (3), such as plan check and QA. Some of the action level results could be resolved to optimal by using MV based kQ in the reference dosimetry as the audit team did.

      4. Discussion

      Results of this audit confirmed and further illustrated dosimetric problems with selected planning system – delivery system combinations. This applied to the modelling of the Elekta wedge in Xio [
      • Lye J.
      • Kenny J.
      • Lehmann J.
      • Dunn L.
      • Kron T.
      • Alves A.
      • et al.
      A 2D ion chamber array audit of wedged and asymmetric fields in an inhomogeneous lung phantom.
      ,
      • Petrovic B.
      • Grzadziel A.
      • Rutonjski L.
      • Slosarek K.
      Linear array measurements of enhanced dynamic wedge and treatment planning system (TPS) calculation for 15 MV photon beam and comparison with electronic portal imaging device (EPID) measurements.
      ] and to the overestimating of the dose behind low density inhomogeneities by Varian’s AAA [
      • Dunn L.
      • Lehmann J.
      • Lye J.
      • Kenny J.
      • Kron T.
      • Alves A.
      • et al.
      National dosimetric audit network finds discrepancies in AAA lung inhomogeneity corrections.
      ,
      • Kry S.F.
      • Alvarez P.
      • Molineu A.
      • Amador C.
      • Galvin J.
      • Followill D.S.
      Algorithms used in heterogeneous dose calculations show systematic differences as measured with the Radiological Physics Center's anthropomorphic thorax phantom used for RTOG credentialing.
      ,
      • Clark C.H.
      • Hurkmans C.W.
      • Kry S.F.
      The role of dosimetry audit in lung SBRT multi-centre clinical trials.
      ]. The latter also extended to beam setups, which resemble a typical breast cancer treatment beam placement (Fig. 5). While there is also a wedge in place, wedge calculation results for AAA have been shown to be accurate (Fig. 6) so the effect is likely due to AAA’s handling of inhomogeneities. The resulting overestimation of the dose to the target was consistent with findings by Petillion et al. who reported a shrinking of the 95% and 100% isodoses in their breast treatment plans. Starting with pencil beam calculations (PBC) the isodoses shrunk when going to AAA, further when calculating with Acuros, dose to medium, and again further when Acuros, dose to water [
      • Petillion S.
      • Swinnen A.
      • Defraene G.
      • Verhoeven K.
      • Weltens C.
      • Van den Heuvel F.
      The photon dose calculation algorithm used in breast radiotherapy has significant impact on the parameters of radiobiological models.
      ]. This in turn agreed with Yoo et al. who described that plans with AAA showed significant underdosage (p = 0.002) of the target volume compared to the original PBC plans. [
      • Yoo S.
      • Wu Q.
      • O'Daniel J.
      • Horton J.
      • Yin F.F.
      Comparison of 3D conformal breast radiation treatment plans using the anisotropic analytical algorithm and pencil beam convolution algorithm.
      ] Hence the target area in a breast cancer treatment with tangential beams likely receives a lower dose than calculated with AAA.
      Measurements inside lung material of the phantom showed a variation in the opposite direction for AAA. Measured dose was found to be higher than calculated. This was more pronounced for the right lateral beam with the measurement point in the left lung. As this beam first traversed the right lung and then water equivalent material in the middle of the phantom before reaching the measurement point in the left lung a combination of effects is assumed. Hence, using AAA for dose calculation risks under-dosing a target beyond lung and over-dosed the lung upstream and downstream of the target. For the Monaco algorithm, which calculated also low inside the lung, yet equally for both lateral beams, the cause is possibly the reported dose being dose to medium. Ion chamber measurements inside the lung carry a larger uncertainty. Therefore these results are RNS at this point.
      Non-optimal results for audited facilities were often a combination of the inherent algorithm difficulties with inhomogeneities or wedges and local TPS implementation. This is in agreement with Kry et al., who wrote that dosimetric errors usually originated with the TPS beam model, and that more focus and attention should be given to beam model commissioning and QA [
      • Kry S.F.
      • Dromgoole L.
      • Alvarez P.
      • Leif J.
      • Molineu A.
      • Taylor P.
      • et al.
      Radiation therapy deficiencies identified during on-site dosimetry visits by the imaging and radiation oncology core Houston quality assurance center.
      ].
      The ACDS had a unique opportunity to implement high level onsite audits with sufficient resources while elsewhere site visits were no longer routinely offered because of budgetary constraints [
      • Kry S.F.
      • Dromgoole L.
      • Alvarez P.
      • Leif J.
      • Molineu A.
      • Taylor P.
      • et al.
      Radiation therapy deficiencies identified during on-site dosimetry visits by the imaging and radiation oncology core Houston quality assurance center.
      ] or audits rely on volunteers [
      • Clark C.H.
      • Aird E.G.
      • Bolton S.
      • Miles E.A.
      • Nisbet A.
      • Snaith J.A.
      • et al.
      Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
      ].
      Results from the presented comprehensive, high accuracy Level III 3D-CRT audit of Australian radiotherapy facilities illustrated dosimetric problems in clinically used dose calculation algorithms mainly related to low density inhomogeneities but also to wedges. Most facilities passed the audit. However, the number of action level and out of tolerance result showed that 3D-CRT dosimetry audits remain relevant and can identify fundamental global and local problems that will also affect advanced treatments.

      Conflict of interest

      None declared.

      Appendix A. Supplementary data

      References

        • Peters L.J.
        • O'Sullivan B.
        • Giralt J.
        • Fitzgerald T.J.
        • Trotti A.
        • Bernier J.
        • et al.
        Critical impact of radiotherapy protocol compliance and quality in the treatment of advanced head and neck cancer: results from TROG 02.02.
        J Clin Oncol. 2010; 28: 2996-3001
        • Kron T.
        • Haworth A.
        • Williams I.
        Dosimetry for audit and clinical trials: challenges and requirements.
        J Phys Conf Ser. 2013; 444: 012014
        • Ibbott G.S.
        • Haworth A.
        • Followill D.S.
        Quality Assurance for Clinical Trials.
        Front Oncol. 2013; 3: 311
        • Palmer A.
        • Mzenda B.
        • Kearton J.
        • Wills R.
        Analysis of regional radiotherapy dosimetry audit data and recommendations for future audits.
        Br J Radiol. 2011; 84: 733-742
        • Pettersen M.N.
        • Aird E.
        • Olsen D.R.
        Quality assurance of dosimetry and the impact on sample size in randomized clinical trials.
        Radiother Oncol. 2008; 86: 195-199
        • Izewska J.
        • Lechner W.
        • Wesolowska P.
        Global availability of dosimetry audits in radiotherapy: The IAEA dosimetry audit networks database.
        Phys Imag Radiat Oncol. 2018; 5: 1-4
        • Clark C.H.
        • Aird E.G.
        • Bolton S.
        • Miles E.A.
        • Nisbet A.
        • Snaith J.A.
        • et al.
        Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
        Br J Radiol. 2015; 88: 20150251
        • Thomas R.A.S.
        • Bolt M.A.
        • Bass G.
        • Nutbrown R.
        • Chen T.
        • Nisbet A.
        • et al.
        Radiotherapy reference dose audit in the United Kingdom by the National Physical Laboratory: 20 years of consistency and improvements.
        Phys Imag Radiat Oncol. 2017; 3: 21-27
        • Clark C.H.
        • Hussein M.
        • Tsang Y.
        • Thomas R.
        • Wilkinson D.
        • Bass G.
        • et al.
        A multi-institutional dosimetry audit of rotational intensity-modulated radiotherapy.
        Radiother Oncol. 2014; 113: 272-278
        • Hussein M.
        • Tsang Y.
        • Thomas R.A.
        • Gouldstone C.
        • Maughan D.
        • Snaith J.A.
        • et al.
        A methodology for dosimetry audit of rotational radiotherapy using a commercial detector array.
        Radiother Oncol. 2013; 108: 78-85
        • Lee J.
        • Mayles H.
        • Baker C.
        • Jafari S.
        • Distefano G.
        • Clark C.
        UK SABR Consortium lung dosimetry audit; Relative dosimetry results.
        Radiother Oncol. 2015; 115: S74-S75
        • Followill D.S.
        • Evans D.R.
        • Cherry C.
        • Molineu A.
        • Fisher G.
        • Hanson W.F.
        • et al.
        Design, development, and implementation of the radiological physics center's pelvis and thorax anthropomorphic quality assurance phantoms.
        Med Phys. 2007; 34: 2070-2076
        • Ibbott G.S.
        • Molineu A.
        • Followill D.S.
        Independent evaluations of IMRT through the use of an anthropomorphic phantom.
        Technol Cancer Research Treatment. 2006; 5: 481-487
        • Ibbott S.I.
        • Followill D.S.
        The Radiological Physics Center: 40 years of vigilance and quality assurance for NCI sponsored clinical trials.
        SSDL Newslett. 2010; : 25-30
        • Kry S.F.
        • Molineu A.
        • Kerns J.R.
        • Faught A.M.
        • Huang J.Y.
        • Pulliam K.B.
        • et al.
        Institutional patient-specific IMRT QA does not predict unacceptable plan delivery.
        Int J Radiat Oncol Biol Phys. 2014; 90: 1195-1201
        • Pulliam K.B.
        • Followill D.
        • Court L.
        • Dong L.
        • Gillin M.
        • Prado K.
        • et al.
        A six-year review of more than 13,000 patient-specific IMRT QA results from 13 different treatment sites.
        J Appl Clin Med Phys. 2014; 15: 4935
        • Kerns J.R.
        • Followill D.S.
        • Lowenstein J.
        • Molineu A.
        • Alvarez P.
        • Taylor P.A.
        • et al.
        Agreement between institutional measurements and treatment planning system calculations for basic dosimetric parameters as measured by the imaging and radiation oncology core-Houston.
        Int J Radiat Oncol Biol Phys. 2016; 95: 1527-1534
        • Kry S.F.
        • Dromgoole L.
        • Alvarez P.
        • Leif J.
        • Molineu A.
        • Taylor P.
        • et al.
        Radiation therapy deficiencies identified during on-site dosimetry visits by the imaging and radiation oncology core Houston quality assurance center.
        Int J Radiat Oncol Biol Phys. 2017; 99: 1094-1100
        • Faught A.M.
        • Kry S.F.
        • Luo D.
        • Molineu A.
        • Bellezza D.
        • Gerber R.L.
        • et al.
        Development of a modified head and neck quality assurance phantom for use in stereotactic radiosurgery trials.
        J Appl Clin Med Phys. 2013; 14: 4313
        • Lehmann J.
        • Dunn L.
        • Lye J.E.
        • Kenny J.W.
        • Alves A.D.
        • Cole A.
        • et al.
        Angular dependence of the response of the nanoDot OSLD system for measurements at depth in clinical megavoltage beams.
        Med Phys. 2014; 41: 061712
        • Kerns J.R.
        • Kry S.F.
        • Sahoo N.
        • Followill D.S.
        • Ibbott G.S.
        Angular dependence of the nanoDot OSL dosimeter.
        Med Phys. 2011; 38: 3955-3962
        • Das I.J.
        Radiochromic film: role and applications in radiation dosimetry.
        CRC Press, Taylor & Francis Group, Boca Raton2018
        • Lye J.
        • Kenny J.
        • Lehmann J.
        • Dunn L.
        • Kron T.
        • Alves A.
        • et al.
        A 2D ion chamber array audit of wedged and asymmetric fields in an inhomogeneous lung phantom.
        Med Phys. 2014; 41: 101712
        • de Prez L.
        • Heukelom S.
        • Jansen B.
        • Jansen W.
        • van de Kamer J.
        • van Klink W.
        • et al.
        An on-site dosimetry audit for high-energy electron beams.
        Phys Imag Radiat Oncol. 2018; 5: 44-51
        • Miri N.
        • Lehmann J.
        • Legge K.
        • Vial P.
        • Greer P.B.
        Virtual EPID standard phantom audit (VESPA) for remote IMRT and VMAT credentialing.
        Phys Med Biol. 2017; 62: 4293-4299
        • Miri N.
        • Lehmann J.
        • Legge K.
        • Zwan B.J.
        • Vial P.
        • Greer P.B.
        Remote dosimetric auditing for intensity modulated radiotherapy: A pilot study.
        Phys Imag Radiat Oncol. 2017; 4: 26-31
        • Ebert M.A.
        • Harrison K.M.
        • Cornes D.
        • Howlett S.J.
        • Joseph D.J.
        • Kron T.
        • et al.
        Comprehensive Australasian multicentre dosimetric intercomparison: issues, logistics and recommendations.
        J Med Imaging Radiat Oncol. 2009; 53: 119-131
        • Kron T.
        • Pham D.
        • Roxby P.
        • Rolfo A.
        • Foroudi F.
        Credentialing of radiotherapy centres for a clinical trial of adaptive radiotherapy for bladder cancer (TROG 10.01).
        Radiother Oncol. 2012; 103: 293-298
        • Tsang Y.
        • Carver A.
        • Groom N.
        • Harris C.
        • Faivre-Finn C.
        • Eaton D.
        A multi-centre dosimetry audit on advanced radiotherapy in lung as part of the Isotoxic IMRT study.
        Phys Imag Radiat Oncol. 2017; 4: 17-21
        • Williams I.
        • Kenny J.
        • Lye J.
        • Lehmann J.
        • Dunn L.
        • Kron T.
        The Australian Clinical Dosimetry Service: a commentary on the first 18 months.
        Australas Phys Eng Sci Med. 2012; 35: 407-411
        • Ibbott G.S.
        • Followill D.S.
        • Molineu H.A.
        • Lowenstein J.R.
        • Alvarez P.E.
        • Roll J.E.
        Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials.
        Int J Radiat Oncol Biol Phys. 2008; 71: S71-S75
        • IAEA. TECDOC 1583
        Commissioning of radiotherapy treatment planning systems: Testing for typical external beam treatment techniques.
        International Atomic Energy Agency, Vienna2008
        • Dunn L.
        • Lye J.
        • Kenny J.
        • Lehmann J.
        • Williams I.
        • Kron T.
        Commissioning of optically stimulated luminescence dosimeters for use in radiotherapy.
        Radiat Meas. 2013; 51–52: 31-39
        • Williams I.
        • Kenny J.
        • Lye J.
        • Lehmann J.
        The Australian Clinical Dosimetry Service: the development and delivery of a national dosimetry audit. World congress on medical physics and biomedical engineering May 26–31, 2012.
        Springer, Beijing, China2013: 1215-1218
        • Lye J.
        • Dunn L.
        • Kenny J.
        • Lehmann J.
        • Kron T.
        • Oliver C.
        • et al.
        Remote auditing of radiotherapy facilities using optically stimulated luminescence dosimeters.
        Med Phys. 2014; 41: 032102
        • Dunn L.
        • Lehmann J.
        • Lye J.
        • Kenny J.
        • Kron T.
        • Alves A.
        • et al.
        National dosimetric audit network finds discrepancies in AAA lung inhomogeneity corrections.
        Phys Med. 2015; 31: 435-441
        • Lehmann J.
        • Kenny J.
        • Lye J.
        • Kron T.
        • Williams I.
        Radiation therapists and Level III audits by the Australian Clinical Dosimetry Service.
        Spectrum. 2012; 19: 14-17
        • IAEA. TRS 398
        Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water.
        International Atomic Energy Agency, Vienna2000
        • Petrovic B.
        • Grzadziel A.
        • Rutonjski L.
        • Slosarek K.
        Linear array measurements of enhanced dynamic wedge and treatment planning system (TPS) calculation for 15 MV photon beam and comparison with electronic portal imaging device (EPID) measurements.
        Radiol Oncol. 2010; 44: 199-206
        • Kry S.F.
        • Alvarez P.
        • Molineu A.
        • Amador C.
        • Galvin J.
        • Followill D.S.
        Algorithms used in heterogeneous dose calculations show systematic differences as measured with the Radiological Physics Center's anthropomorphic thorax phantom used for RTOG credentialing.
        Int J Radiat Oncol Biol Phys. 2013; 85: e95-e100
        • Clark C.H.
        • Hurkmans C.W.
        • Kry S.F.
        The role of dosimetry audit in lung SBRT multi-centre clinical trials.
        Phys Med. 2017; 44: 171-176
        • Petillion S.
        • Swinnen A.
        • Defraene G.
        • Verhoeven K.
        • Weltens C.
        • Van den Heuvel F.
        The photon dose calculation algorithm used in breast radiotherapy has significant impact on the parameters of radiobiological models.
        J Appl Clin Med Phys. 2014; 15: 259-269
        • Yoo S.
        • Wu Q.
        • O'Daniel J.
        • Horton J.
        • Yin F.F.
        Comparison of 3D conformal breast radiation treatment plans using the anisotropic analytical algorithm and pencil beam convolution algorithm.
        Radiother Oncol. 2012; 103: 172-177