An audit methodology was developed and applied for output factor (OF) calculations in radiotherapy. The auditees were asked to calculate OFs for field sizes from 10 × 10 cm2 to 2 × 2 cm2. Sixty five beams were audited; missing reference OFs were interpolated. The calculated OFs were in 73% of cases higher than the reference data. The smaller the field size, the higher the overestimations which were observed in the higher fraction of cases. Treatment planning systems generally overestimated OFs for small fields. The reference dataset helped radiotherapy centres to identify discrepancies which were higher than typical.
Appropriate calculation of output factors for small fields shaped by a multileaf collimator (MLC), performed in treatment planning systems (TPS), is essential for intensity-modulated radiation therapy (IMRT). Therefore, proper configuration of beam data and precise modelling of the MLC in the treatment planning system (TPS) are key factors that have to be verified prior to clinical use. Dosimetry for fields smaller than 3 × 3 cm2 is very difficult and has a high degree of uncertainty. This is caused both by the relatively large penumbra size, as well as by the changes in the energy spectrum [
Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams.
]. The Radiological Physics Center (RPC) at the MD Anderson Cancer Center (presently IROC-Houston QA Center) has prepared a set of data containing output factors depending on nominal beam energy, field sizes and accelerator models [
]. However, the published data do not cover all beam energies from the mega-voltage range used in photon-based radiotherapy. The RPC dataset and the interpolation functions proposed in this work were used to carry out a nationwide audit of small field OF calculations.
2. Material and methods
Participants were asked to calculate the output factors for beams formed by the multi-leaf collimator (MLC), using their planning software. The results of their calculations were compared with the reference published data. All 35 Polish radiotherapy departments were invited to take part in the study, and 32 of them responded and provided their results. The TPS calculations for medical accelerators of three vendors were evaluated: Elekta, Siemens and Varian, further denoted throughout the text as type A, B and C. In total, 65 beams were audited: 20 of type A, 15 of type B and 30 of type C accelerators. In seven centres the calculations were repeated for the same beams with two or three different TPSs or with alternative calculation algorithms. In total, 76 beam&TPS combinations were evaluated. Most of the results (90%) were obtained for beams with a nominal energy of either 6 MV (62%) or 15 MV (28%) (see Supplementary Tables S1–S3). The most commonly used beam energies (excluding FFF) for each of the vendors were used in the calculations. The participants had to calculate the number of monitor units (MU) for the delivery a dose of 10 Gy to water with five square, MLC-shaped fields (10 × 10 cm2, 6 × 6 cm2, 4 × 4 cm2, 3 × 3 cm2 and 2 × 2 cm2), to a reference point at a depth of 10 cm on the central axis at a source-to-phantom distance (SPD) of 100 cm. The dose rates DR(f,E)[Gy/MU] were calculated for a specific field size f [cm2] and beam energy E [MV], and then divided by the calculated for a field size of 10 × 10 cm2 and for the same beam energy, thus providing a normalized output factor OF(f,E) (see Eq. (1)).
(1)
The discrepancies between the reference RPC data and the institution OF were analysed and compared with the criteria of acceptability provided by other authors [
TECDOC-1583 Commissioning of radiotherapy treatment planning systems: testing for typical external beam treatment techniques: report of the Coordinated Research Project on Development of Procedures for Quality Assurance of Dosimetry Calculations in Radiotherapy. Vienna: IAEA; 2008.
]. Thus, when the 3% level of disagreement was exceeded, the authors expected institutions to consider this as problematic. The values of the OFs for beam energies E [MV], not present in the RPC data set, have been interpolated with a second degree polynomial (see Eq. (2))
(2)
using the non-linear least-squares (NLLS) Marquardt-Levenberg algorithm [
]. Interpolated values of the reference OFs had to be used in 18% of cases. The clinically used TPSs and related calculation algorithms were examined (see Table 1).
Table 1Treatment planning systems and related calculation algorithms used for the OF calculations.
The audit results for individual participants were grouped for accelerator types A, B and C.
3. Results
The a(f), b(f) and c(f) parameters of the Eq. (2) were obtained in the procedure of fitting to the experimental data (see Supplementary Table S4 and Supplementary Figs. S1–S3). In any case of fitting, the final sum of the square residuals (WSSR) was not larger than . The OFs calculated with TPSs were in 73% of cases higher than the published reference data. The smaller the field size, the higher the overestimations which were observed. Overestimations of OFs were observed in 69% calculations for 6 × 6 cm2 fields, in 70% for 4 × 4 cm2, in 75% for 3 × 3 cm2, and in 77% for 2 × 2 cm2 fields. The mean values (±σ) of the fraction of OFs calculated with TPS to the reference data were: 1.001 (±0.007) for 6 × 6 cm2, 1.004 (±0.010) for 4 × 4 cm2 1.008 (±0.012) for 4 × 4 cm2, and 1.014 (±0.024) for 2 × 2 cm2 fields (see Supplementary Figs. S4–S7). For smaller field sizes, wider distributions and higher modal values of ratios of the institution OF to the reference OF were observed (see Fig. 1). The ratios of the audited institution mean OFs to the corresponding values reported by RPC, were generally within a range of 0.996–1.002. Only for type B accelerator 6 MV and field size 2 × 2 cm2, this ratio was as high as 1.04. For type A accelerators, all calculation results showed a deviation from the reference values lower than 3%. For type B and C accelerators, the resulting calculations for fields larger than 2 × 2 cm2 differed by less than 4%. For 2 × 2 cm2 fields formed by MLCs of B and C linacs, the differences between the calculated and measured output factors often exceeded 5% and were below 10%.
Fig. 1Distribution of the audit results for different field sizes in terms of ratios of output factors obtained from audited institutions (OFTPS) to the reference data (OFRPC) including interpolated values. On the vertical axis are the cardinalities of the results grouped in 0.01 wide classes, presented on the horizontal axis.
This audit of small field calculation of OFs was performed in a significant number of institutions for a wide range of linac models and TPS. As reference data, OFs measured and published by other authors were used [
]. Interpolated OFs were generated for missing reference data. The universal formula derived from interpolation for the measurement data, was proposed here. This allowed calculation of the expected reference OF for most currently used models of accelerators for a wide range of nominal energies for beams modified with flattening filters.
The mean values (see Table 2) and the standard deviations of calculated relative output factors obtained in the audit were similar to the RPC values. Average absolute deviations from the mean values of calculated (institution) OFs acquired by RPC were 1.1%, 1.4% and 1.3% for A, B and C accelerators, respectively. The standard deviations of the calculated OFs were ≤0.04 in both audits. Previous results of Monte Carlo simulations [
Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams.
] were extensively exploited by the RPC in audits. The standard deviations of the RPC results are below 2.4%, which is comparable with the recommendations of 2% tolerance in comparisons between TPS-calculated and measured OFs for field sizes of 3 × 3 cm2 and larger [
TECDOC-1583 Commissioning of radiotherapy treatment planning systems: testing for typical external beam treatment techniques: report of the Coordinated Research Project on Development of Procedures for Quality Assurance of Dosimetry Calculations in Radiotherapy. Vienna: IAEA; 2008.
]. The average standard deviations of measured OF within groups of the same accelerator type for different beam energies and field sizes were 0.9%, 0.7% and 0.6% for type A, B and C accelerator, respectively [
]. This means that the construction differences among classes available within the same accelerator type are minor and do not significantly influence the beam or the collimation sections. We consider that it is justified to refer to the published results of measurements performed for a large number of accelerators before referring to measurements performed for each specific unit installed at the audited institutions, because the OF measurements for TPS configurations are typically performed for field sizes ≥3 × 3 cm2, and because precise small field dosimetry in less advanced institutions is still limited. In the case of very small fields, the positioning or correction factor uncertainties would make the audit results confusing, if the measurements were carried out in different institutions with different equipment. The audit intended to highlight that institutions should compare their beam data and the TPS calculations also with the reference data. All audited institutions reported the commissioning measurements performed for small field sizes (as close as possible to 4 × 4 cm2). This is consistent with IAEA guidance [
TECDOC-1583 Commissioning of radiotherapy treatment planning systems: testing for typical external beam treatment techniques: report of the Coordinated Research Project on Development of Procedures for Quality Assurance of Dosimetry Calculations in Radiotherapy. Vienna: IAEA; 2008.
Table 2Mean values of output factors calculated with TPSs for beams of type A, B and C accelerators. The values in parentheses below are the standard deviations of collected TPS data. The values in square brackets are the average absolute percent differences between reference data and TPS data. For each beam type and field size, the cardinality n of TPS calculated results is shown.
The aim of this study was not to repeat the whole on-site audit procedure performed by the RPC and, in our opinion, there is no need to repeat the measurements in future routine national audit programmes unless unresolved high discrepancies are observed by the auditee. The simple interpolation formula presented by the authors of this work needs further validation through carrying out and juxtaposing additional measurements and by means of Monte Carlo simulations. Similar work has been performed for electron and photon beams [
]. Reports presenting the ratios of the OF to the reference values calculated in institutions were sent back to the participants. Recalculated OFs showing improved results for five beams were obtained from four centres. In the first center (A/MO), the auditee performed new measurements and changed the beam modelling in the most advanced TPS used in the institution. In the second institution (C/EC), the PB calculation algorithm was replaced with an AA one. In the third institution (A/PC), the second alternative “small” beam model was created for the same beam, which was optimized and used only for field sizes from 1.5 × 1.5 cm2 up to 10 × 10 cm2. In the fourth institution, misunderstanding of the geometrical set-up from the audit instruction was reported. In one institution (B/PR), the OF factor value for field size 2 × 2 cm2 was set in TPS, as the same as for the 3 × 3 cm2 field. The institution did not change the TPS configuration, because 2 × 2 cm2 fields were not clinically used there. The results presented here include the corrected values of the OFs. It seems that there is a correlation between the results of the OF calculations and the type of the TPS calculation algorithm. For type B accelerators and ON/PB TPS, all results for the 2 × 2 cm2 field differ more than 5% from the reference data. For type C accelerators, upgrading TPS from EC/PB to EC/AA decreased the difference for the 2 × 2 cm2 field from 8.5% to 2.2%, relative to the reference data. The authors believe that the reference dataset is an easily available data source which can help users of linacs equipped with MLCs to detect potential problems in IMRT delivery. The audit aims were to verify the quality of the TPS calculations in the case of IMRT-style small fields, using the postal method, and to popularize the method of TPS QA, which does not require time-consuming and difficult measurements. In general, Eq. (2) can probably be used also for FFF beams, but the coefficients presented in the Supplementary Table S4 are valid only for filtered beams.
The set of measured small field output factors provided by the RPC, together with the parameters of the analytical functions published here, is a very good tool for TPS QA. Follow-up actions have been performed in the four audited centres with issues, yielding improved results. This audit already has had substantial impact for clinical practice but more reference data sets, which are anticipated for small field OFs for FFF beams, could also be used in a future audit of this type.
Conflict of Interest
Krzysztof Chełmiński and Wojciech Bulski have no conflict of interest to declare.
Acknowledgements
The methodology of the audit of small field output performance was established within the framework of the CRP E2.40.16 project “Development of Quality Audits for Radiotherapy Dosimetry for Complex Treatment Techniques”, run by the Health Section of the International Atomic Energy Agency (IAEA).
Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams.
TECDOC-1583 Commissioning of radiotherapy treatment planning systems: testing for typical external beam treatment techniques: report of the Coordinated Research Project on Development of Procedures for Quality Assurance of Dosimetry Calculations in Radiotherapy. Vienna: IAEA; 2008.